FDA Adverse Event Injury Summary report: N

FINELINE II

MDR report key: 2913204 · Received January 14, 2013

Report

Report Number
2124215-2012-17252
Event Type
Injury
Date Received
January 14, 2013
Date of Event
December 17, 2012
Report Date
June 13, 2013
Manufacturer
CPI - DEL CARIBE
Product Code
NVN
PMA / PMN Number
P960004
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IN
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

AS NO FURTHER INFORMATION CONCERNING THIS REPORT IS EXPECTED, OUR INVESTIGATION IS COMPLETE. THIS INVESTIGATION WILL BE UPDATED SHOULD FURTHER INFORMATION BE PROVIDED.

Additional Manufacturer Narrative · 1

UPON RECEIPT AT OUR POST MARKET QUALITY ASSURANCE LABORATORY VISUAL INSPECTION NOTED THAT ONLY THE PROXIMAL SEGMENT OF THE LEAD WAS RETURNED, AND THE INSULATION WAS TWISTED INDICATING TWIDDLERS SYNDROME. THERE WAS NOTED ELECTROCAUTERY DAMAGE ON THE LEAD. THE DAMAGE NOTED TO THE LEAD WAS DETERMINED TO BE AS A RESULT OF TWIDDLERS SYNDROME.

Description of Event or Problem · 1

--

Description of Event or Problem · 1

BOSTON SCIENTIFIC RECEIVED INFORMATION THAT THIS PATIENT'S RIGHT VENTRICULAR (RV) LEAD WAS EXHIBITING LOSS OF CAPTURE AND INCREASED THRESHOLDS. THE RIGHT ATRIAL (RA) LEAD WAS ALSO EXHIBITING ISSUES OF INCREASED IMPEDANCES. A CHEST X-RAY WAS TAKEN WHICH INDICATED THAT THE PATIENT WAS A TWIDDLER, AND TWIDDLED THE LEADS AROUND THE DEVICE. A REVISION PROCEDURE WAS DONE, WHICH SURGICALLY ABANDONED BOTH LEADS, AND TWO NEW LEADS WERE THEN IMPLANTED. POST IMPLANT, THERE WAS ONCE AGAIN LOSS OF CAPTURE ON THE NEWLY IMPLANTED RV LEAD. THEY PHYSICIAN THOUGHT THAT MAYBE THE DEVICE HEADER WAS DAMAGED FROM THE TWIDDLING, SO HE ELECTED TO EXPLANT AND REPLACE THE DEVICE. TO DATE, NO ADDITIONAL ADVERSE PATIENT EFFECTS HAVE BEEN REPORTED.