TELIGEN
Report
- Report Number
- 2124215-2012-17135
- Event Type
- Malfunction
- Date Received
- January 14, 2013
- Date of Event
- September 10, 2012
- Report Date
- February 27, 2013
- Manufacturer
- GUIDANT CRM CLONMEL IRELAND
- Product Code
- LWS
- PMA / PMN Number
- P010012
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- OTHER
Narratives
THE PROGRAMMED CONFIGURATION WAS REPROGRAMMED TO RV COIL TO CAN AND ACCEPTABLE SHOCK IMPEDANCE MEASUREMENTS WERE OBSERVED. THE PHYSICIAN WILL CONTINUE TO MONITOR THE DEVICE AND LEAD AND CONSIDER DEFIBRILLATION THRESHOLD (DFT) TESTING IN THE FUTURE. AS NO FURTHER INFORMATION CONCERNING THIS REPORT IS EXPECTED, OUR INVESTIGATION IS COMPLETE. THIS INVESTIGATION WILL BE UPDATED SHOULD FURTHER INFORMATION BE PROVIDED.
SUBSEQUENT INFORMATION WAS RECEIVED THAT THIS PRODUCT WAS EXPLANTED AND REPLACED TWO MONTHS LATER. UPON RECEIPT AT OUR POST MARKET QUALITY ASSURANCE LABORATORY, A THOROUGH EVALUATION OF THE DEVICE WAS PERFORMED. A VISUAL INSPECTION OF THE DEVICE HEADER AND CASE NOTED NO ANOMALIES. THE DEVICE WAS THEN EXPOSED TO SIMULATED HEART LOAD CONDITIONS, AND THE DEFIBRILLATION, PACING AND SENSING FUNCTIONS WERE TESTED. THE DEVICE OPERATED APPROPRIATELY, ACCORDING TO ITS PERFORMANCE SPECIFICATIONS WITH NO INTERRUPTIONS IN THERAPY OUTPUT OR PROGRAMMER COMMUNICATION AT THE RETURNED PROGRAMMED SETTINGS. A SERIES OF ELECTRICAL TESTS WAS ALSO PERFORMED, AND AGAIN, NORMAL DEVICE FUNCTION WAS OBSERVED. LABORATORY ANALYSIS DID NOT IDENTIFY ANY DEVICE CHARACTERISTICS THAT WOULD HAVE CAUSED OR CONTRIBUTED TO THE REPORTED CLINICAL OBSERVATIONS.
ADDITIONAL INFORMATION WAS RECEIVED FROM THE LOCAL FIELD REPRESENTATIVE THAT AFTER THE SHOCKING CONFIGURATION HAD BEEN REPROGRAMMED, DEFIBRILLATION THRESHOLD (DFT) TESTING WAS COMPLETED SUCCESSFULLY AND REVEALED STABLE MEASUREMENTS. AS NO FURTHER INFORMATION CONCERNING THIS REPORT IS EXPECTED, OUR INVESTIGATION IS COMPLETE. THIS INVESTIGATION WILL BE UPDATED SHOULD FURTHER INFORMATION BE PROVIDED.
BOSTON SCIENTIFIC RECEIVED INFORMATION THAT THIS IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD) AND IMPLANTABLE DEFIBRILLATION LEAD EXHIBITED LOW OUT OF RANGE SHOCK IMPEDANCE MEASUREMENTS THAT WERE ABLE TO BE REPRODUCED WITH POCKET MANIPULATION. IT WAS NOTED THAT THE SHOCK VECTOR HAD BEEN PROGRAMMED TRIAD. BOSTON SCIENTIFIC TECHNICAL SERVICES (TS) DISCUSSED THE POSSIBILITY OF A LEAD INSULATION ISSUE VERSUS A CONNECTION ISSUE AND DISCUSSED TROUBLESHOOTING OPTIONS. NO ADVERSE PATIENT EFFECTS WERE REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 19742 | TELIGEN | IMPLANTABLE CARDIOVERTER DEFIBRILLATOR | LWS | GUIDANT CRM CLONMEL IRELAND | E102 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 43 YR | Hospitalization| R | E102| 0184| 4136| T175 |