FDA Adverse Event Malfunction Summary report: N

PROGRAMMING SOFTWARE

MDR report key: 2913023 · Received January 14, 2013

Report

Report Number
1644487-2013-00139
Event Type
Malfunction
Date Received
January 14, 2013
Date of Event
December 17, 2012
Report Date
December 17, 2012
Manufacturer
CYBERONICS INC
Product Code
LYJ
PMA / PMN Number
P970003
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OH, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

.

Description of Event or Problem · 1

A VNS TREATING NEUROLOGIST REPORTED THAT THE DELL X50 HANDHELD COMPUTER WAS NOT TURN ON. THE COMPANY REPRESENTATIVE FOLLOWED UP WITH THE NEUROLOGIST'S OFFICE AND PERFORMED TROUBLESHOOTING ON THE COMPUTER. THE DEVICE WAS ABLE TO BE TURNED ON AFTER PERFORMING A HARD RESET. NO ADDITIONAL INFORMATION WAS PROVIDED BY THE PHYSICIAN. THE SPECIFIC PRODUCT INFORMATION IS UNKNOWN, AS THE PHYSICIAN HAS TWO DELL X50 HANDHELD COMPUTERS AND THE COMPUTER THAT COULD NOT BE TURNED ON WAS NOT SPECIFIED. HOWEVER, BOTH PROGRAMMING SYSTEMS WERE FUNCTIONING PROPERLY AFTER THE HARD RESET.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
20169 PROGRAMMING SOFTWARE PROGRAMMING COMPUTER LYJ CYBERONICS INC MODEL 250 NI

Patients

Seq Age Sex Outcome Treatment
1