FDA Adverse Event Injury Summary report: N

JAWS INSERT, 330 MM WORKING LENGTH

MDR report key: 2912912 · Received January 9, 2013

Report

Report Number
9610773-2012-00241
Event Type
Injury
Date Received
January 9, 2013
Date of Event
December 4, 2012
Report Date
December 28, 2012
Manufacturer
OLYMPUS WINTER & IBE GMBH
Product Code
FCL
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
NURSE

Narratives

Additional Manufacturer Narrative · 1

THE SUSPECT MEDICAL DEVICE WAS NOT RETURNED TO THE MFR FOR EVAL BUT TO THE OLYMPUS DISTRIBUTOR (B)(6). AN INVESTIGATION OUTCOME IS NOT AVAILABLE YET. THEREFORE, THE EXACT CAUSE OF THE DEFECT (FALLEN OFF TIP) CANNOT BE CONCLUSIVELY DETERMINED AT THIS TIME. THERE WAS NO REPORT ABOUT PT INJURY AND THE FRAGMENT COULD BE RETRIEVED FROM THE PT SUCCESSFULLY. THE PROCEDURE WAS COMPLETED. THIS REPORT IS BEING SUBMITTED AS A MEDICAL DEVICE REPORT IN ABUNDANCE OF CAUTION.

Description of Event or Problem · 1

DURING THE PROCEDURE OF LAPAROSCOPIC APPENDECTOMY, THE PHYSICIAN TRIED TO GRASP THE PT'S TISSUE WITH THE SUSPECT MEDICAL DEVICE (JAWS INSERT, 330 MM WORKING LENGTH, CROCEOLMI GRASPING FORCEPS) BUT COULDN'T MANAGE IT. THE PHYSICIAN WITHDREW THE DEVICE FROM THE PT AND CONFIRMED THAT THE TIP OF THE GRASPING FORCEPS WAS MISSING. HE TOOK AN X-RAY TO CHECK THE FRAGMENT, AND RETRIEVED IT FROM THE PT'S CAVITY (SIZE OF THE FRAGMENT: APPROX. 5 MM). JUDGING FROM THE MAIN BODY OF THE DEVICE AND THE RETRIEVED FRAGMENT, THE PHYSICIAN DETERMINED THAT NO FRAGMENT REMAINED INSIDE THE PT ANYMORE. THEN THE PHYSICIAN TOOK AN X-RAY AGAIN TO CONFIRM THAT NO FRAGMENT REMAINED INSIDE THE PT, JUST IN CASE. THE PHYSICIAN RE-STARTED THE PROCEDURE USING OTHER FORCEPS AND THE PROCEDURE WAS COMPLETED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
12121 JAWS INSERT, 330 MM WORKING LENGTH CROCEOLMI GRASPING FORCEPS FCL OLYMPUS WINTER & IBE GMBH 123W-0049

Patients

Seq Age Sex Outcome Treatment
1 15 YR Other| R