FDA Adverse Event
Malfunction
Summary report: N
LIFEVEST WCD 4000 SYSTEM
MDR report key: 2912854
·
Received November 30, 2012
Report
- Report Number
- 3008642652-2012-03145
- Event Type
- Malfunction
- Date Received
- November 30, 2012
- Date of Event
- November 16, 2012
- Report Date
- November 27, 2012
- Manufacturer
- ZOLL LIFECOR CORPORATION
- Product Code
- MVK
- PMA / PMN Number
- P010030
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- PA, US
- Reporter Occupation
- BIOMEDICAL ENGINEER
Narratives
Additional Manufacturer Narrative · 1
DEVICE EVALUATION OF ELECTRODE BELT SN (B)(4) HAS BEEN COMPLETED. UPON EVALUATION THE PULSE WIRE IN THE REAR THERAPY ELECTRODE INTERCONNECT CABLE HAD AN INTERMITTENT OPEN CONNECTION. THE ROOT CAUSE FOR THE OPEN PULSE WIRE COULD NOT BE POSITIVELY IDENTIFIED BUT WAS LIKELY EXCESSIVE FORCE ON THE CABLE SECTION. NO ADVERSE EVENT RESULTED FROM THE DEFECTIVE ELECTRODE BELT. THE LAST PT TO USE THIS MONITOR DID NOT REPORT ANY DEFICIENCIES.
Description of Event or Problem · 1
A REVIEW OF SERVICE DATA DETECTED A REPORTABLE PROBLEM. DURING SERVICING, ELECTRODE BELT SN (B)(4) FAILED A THERAPY ELECTRODE RECOGNITION TEST. THE LAST PT TO USE THIS ELECTRODE BELT DID NOT REPORT ANY DEFICIENCIES.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | LIFEVEST WCD 4000 SYSTEM | WEARABLE CARDIOVERTER DEFIBRILLATOR | MVK | ZOLL LIFECOR CORPORATION | WCD 4000 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA |