FDA Adverse Event Injury Summary report: N

ON-Q PAIN PUMP

MDR report key: 2912764 · Received January 8, 2013

Report

Report Number
2026095-2012-00415
Event Type
Injury
Date Received
January 8, 2013
Date of Event
May 10, 2007
Report Date
December 11, 2012
Manufacturer
I-FLOW, LLC
Product Code
MEB
PMA / PMN Number
K063530
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

METHOD: NO PRODUCT WAS RETURNED FOR EVALUATION AND INVESTIGATION. RESULTS: THE INFORMATION CONTAINED IS FROM LEGAL DOCUMENTS SERVED ON I-FLOW, LLC. THE COMPLAINT WAS OPENED SO THAT A MEDICAL DEVICE REPORT (MDR) CAN BE FILED WITH THE US FOOD AND DRUG ADMINISTRATION. NO FURTHER INVESTIGATION WILL BE CONDUCTED. CONCLUSION: AS THIS COMPLAINT WAS CREATED FROM A LAWSUIT SERVED ON I-FLOW OR THE THREAT OF A LAWSUIT, NO CUSTOMER CONTACT CAN BE MADE AT THIS TIME DUE TO THE PENDING OR THREATENED LITIGATION." AS OF 11/09/2006 I-FLOW UPDATED THE ON-Q PUMP DIRECTIONS FOR USE (DFU), TO INCLUDE THE FOLLOWING IN THE WARNINGS SECTION: "AVOID PLACING THE CATHETER IN JOINT SPACES. ALTHOUGH THERE IS NO DEFINITIVE ESTABLISHED CAUSAL RELATIONSHIP, SOME LITERATURE HAS SHOWN A POSSIBLE ASSOCIATION BETWEEN CONTINUOUS INTRA-ARTICULAR INFUSIONS (PARTICULARLY WITH BUPIVACAINE) AND THE SUBSEQUENT DEVELOPMENT OF CHONDROLYSIS." (DFU 1307011). ON 08/08/2007, I-FLOW HAS ALSO PREPARED A TECHNICAL BULLETIN ENTITLED "WHAT WE KNOW ABOUT CHONDROLYSIS TODAY". (1303722, REV. E).

Description of Event or Problem · 1

DRUG/DILUENT: MARCAINE 0.5%; FILL VOLUME: 100ML; FLOW RATE: 2ML/HR; PROCEDURE: OPEN REDUCTION AND INTERNAL FIXATION (ORIF) OF THE GREATER TUBEROSITY FRACTURE (LEFT). ROTATOR CUFF REPAIR. CATHPLACE: SUBACROMIAL SPACE. REFERENCE: 2026095-2012-00416 ((B)(6)). PATIENT ALLEGES CARTILAGE DAMAGE IN LEFT SHOULDER FOLLOWING PLACEMENT OF AN I-FLOW ON-Q PAINBUSTER, PM012, AFTER SURGERY ON (B)(6) 2007. A SUBSEQUENT SURGERY WAS PERFORMED ON THE LEFT SHOULDER ON (B)(6) 2009.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
9811 ON-Q PAIN PUMP ELASTOMERIC PUMP MEB I-FLOW, LLC PM012 6A2508

Patients

Seq Age Sex Outcome Treatment
1 29 YR Other