FDA Adverse Event Injury Summary report: N

ON-Q PAIN PUMP

MDR report key: 2912715 · Received January 8, 2013

Report

Report Number
2026095-2012-00344
Event Type
Injury
Date Received
January 8, 2013
Date of Event
December 8, 2005
Report Date
December 11, 2012
Manufacturer
I-FLOW, LLC
Product Code
MEB
PMA / PMN Number
K063530
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

METHOD: NO PRODUCT WAS RETURNED FOR EVALUATION AND INVESTIGATION. THE LOT NUMBER WAS NOT PROVIDED; THEREFORE, THE DEVICE HISTORY RECORDS COULD NOT BE REVIEWED. RESULTS: THE INFORMATION CONTAINED IS FROM LEGAL DOCUMENTS SERVED ON I-FLOW, LLC. THE COMPLAINT WAS OPENED SO THAT A MEDICAL DEVICE REPORT (MDR) CAN BE FILED WITH THE US FOOD AND DRUG ADMINISTRATION. NO FURTHER INVESTIGATION WILL BE CONDUCTED. CONCLUSION: AS THIS COMPLAINT WAS CREATED FROM A LAWSUIT SERVED ON I-FLOW OR THE THREAT OF A LAWSUIT, NO CUSTOMER CONTACT CAN BE MADE AT THIS TIME DUE TO THE PENDING OR THREATENED LITIGATION." AS OF 11/09/2006, I-FLOW UPDATED THE ON-Q PUMP DIRECTIONS FOR USE (DFU), TO INCLUDE THE FOLLOWING IN THE WARNINGS SECTION: "AVOID PLACING THE CATHETER IN JOINT SPACES. ALTHOUGH THERE IS NO DEFINITIVE ESTABLISHED CAUSAL RELATIONSHIP, SOME LITERATURE HAS SHOWN A POSSIBLE ASSOCIATION BETWEEN CONTINUOUS INTRA-ARTICULAR INFUSIONS (PARTICULARLY WITH BUPIVACAINE) AND THE SUBSEQUENT DEVELOPMENT OF CHONDROLYSIS." (DFU 1307011). ON 08/08/2007, I-FLOW HAS ALSO PREPARED A TECHNICAL BULLETIN ENTITLED "WHAT WE KNOW ABOUT CHONDROLYSIS TODAY." (NUMBER 1303722, REV. E).

Description of Event or Problem · 1

DRUG/DILUENT: MARCAINE 0.25%. FILL VOLUME: 120ML. FLOW RATE: UNKNOWN. PROCEDURE: ARTHROSCOPIC ROTATOR CUFF DECOMPRESSION RIGHT SHOULDER WITH EXCISION OF DISTAL CLAVICLE AND ROTATOR CUFF REPAIR. CATHPLACE: SUBACROMIAL SPACE. REFERENCE: 2026095-2012-00422 ((B)(4)). PATIENT ALLEGES CARTILAGE DAMAGE IN RIGHT SHOULDER FOLLOWING PLACEMENT OF AN I-FLOW ON-Q PAINBUSTER, PM012, AFTER SURGERY ON (B)(6) 2005. PATIENT HAD SURGERY ON THE LEFT SHOULDER ON (B)(6) 2006.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
8637 ON-Q PAIN PUMP ELASTOMERIC PUMP MEB I-FLOW, LLC PM012 UNKNOWN

Patients

Seq Age Sex Outcome Treatment
1 48 YR Other