FDA Adverse Event Malfunction Summary report: N

FENESTRATED BIPOLAR FORCEPS INSTRUMENT

MDR report key: 2912366 · Received January 11, 2013

Report

Report Number
2955842-2013-00159
Event Type
Malfunction
Date Received
January 11, 2013
Date of Event
December 4, 2012
Report Date
December 13, 2012
Manufacturer
INTUITIVE SURGICAL,INC.
Product Code
NAY
PMA / PMN Number
K050369
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NY
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE INSTRUMENT WAS RETURNED AND EVALUATED. COMPLAINT MISALIGNED IS CONFIRMED. ONE GRIP IS BENT, CAUSING SIDE TO SIDE MISALIGNMENT OF GRIPS. THERE IS A .062 OFFSET AT THE TIPS, INDICATING OVERLOADING AT THE TIP. ELECTRICAL CONTINUITY PASSED. ADDITIONAL FINDING: FRAYED PITCH CABLE. INSTRUMENT WAS FOUND WITH FRAYED PITCH CABLE AT DISTAL CLEVIS HUB. NO DAMAGE WAS FOUND AT THE CLEVIS. NO OTHER CABLE DAMAGE WAS FOUND. LAST FINDING SCRATCHES ON MAINTUBE. DISTAL END OF MAIN TUBE HAS VARIOUS SCRATCH MARKS WITH LIGHT MATERIAL REMOVAL. SUGGESTING THE MAY HAVE BEEN CAUSED BY INSTRUMENT COLLISIONS OR ROUGH HANDLING. THE CUSTOMER REPORTED COMPLAINT DOES NOT ITSELF CONSTITUTE A MDR REPORTABLE EVENT; HOWEVER THE REPORTED MALFUNCTION IF TO RECUR COULD CAUSE OR CONTRIBUTE TO AN ADVERSE EVENT. THE INSTRUMENTS AND ACCESSORIES USER MANUAL SPECIFICALLY STATES: GENERAL PRECAUTIONS AND WARNINGS - HANDLE INSTRUMENTS WITH CARE. AVOID MECHANICAL SHOCK OR STRESS THAT CAN CAUSE DAMAGE TO THE INSTRUMENTS. DO NOT USE AN INSTRUMENT TO CLEAN DEBRIS FROM ANOTHER INSTRUMENT INTRAOPERATIVELY. THIS MAY RESULT IN DAMAGE TO THE INSTRUMENTS OR OTHER UNINTENDED CONSEQUENCES, SUCH AS DISCONNECTION OF THE INSTRUMENT TIP.

Description of Event or Problem · 1

IT WAS REPORTED THAT PRIOR TO STARTING A DA VINCI SI HYSTERECTOMY PROCEDURE, THE SURGICAL STAFF REPORTED SEEING JAWS OF THE FENESTRATED BIPOLAR FORCEPS INSTRUMENT WERE MISALIGNED. NO PATIENT HARM, ADVERSE OUTCOME OR INJURY WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
19132 FENESTRATED BIPOLAR FORCEPS INSTRUMENT ELECTROSURGICAL INSTRUMENT NAY INTUITIVE SURGICAL,INC. 420205-05 M10121018 447

Patients

Seq Age Sex Outcome Treatment
1 44 YR DA VINCI SI SYSTEM INSTRUMENTS AND ACCESSORIES