PK DISSECTING FORCEPS INSTRUMENT
Report
- Report Number
- 2955842-2013-00161
- Event Type
- Malfunction
- Date Received
- January 11, 2013
- Date of Event
- December 10, 2012
- Report Date
- December 13, 2012
- Manufacturer
- INTUITIVE SURGICAL,INC.
- Product Code
- NAY
- PMA / PMN Number
- K061260
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA
- Reporter Occupation
- OTHER
Narratives
THE INSTRUMENT WAS RETURNED AND EVALUATED. THE CUSTOMER INITIALLY REPORTED FRAYED WIRE, HOWEVER, FOR CLARIFICATION THE DAMAGED INSTRUMENT COMPONENT WAS A BROKEN PITCH CABLE. BASED ON THIS CLARIFICATION, THE CUSTOMER REPORTED COMPLAINT IS CONFIRMED. THE PITCH CABLE IS BROKEN AT THE DISTAL CLEVIS HUB. CABLE SEGMENT THAT CONTAINS THE CRIMP IS STILL INSTALLED IN CLEVIS. CLEVIS DOES NOT EXHIBIT ANY DAMAGE OR WEAR MARKS. ELECTRICAL CONTINUITY PASSED. ADDITIONAL FINDING: SCRATCHES ON MAINTUBE. DISTAL END OF MAIN TUBE HAS VARIOUS SCRATCH MARKS WITH LIGHT MATERIAL REMOVAL. SCRATCHES ARE SHORT IN LENGTH AND NOT ALIGNED WITH THE TUBE AXIS, SUGGESTING THE MAY HAVE BEEN CAUSED BY INSTRUMENT COLLISIONS OR ROUGH HANDLING. THE INSTRUMENTS AND ACCESSORIES USER MANUAL SPECIFICALLY STATES: GENERAL PRECAUTIONS AND WARNINGS - HANDLE INSTRUMENTS WITH CARE. AVOID MECHANICAL SHOCK OR STRESS THAT CAN CAUSE DAMAGE TO THE INSTRUMENTS. DO NOT USE AN INSTRUMENT TO CLEAN DEBRIS FROM ANOTHER INSTRUMENT INTRAOPERATIVELY. THIS MAY RESULT IN DAMAGE TO THE INSTRUMENTS OR OTHER UNINTENDED CONSEQUENCES, SUCH AS DISCONNECTION OF THE INSTRUMENT TIP.
IT WAS REPORTED THAT DURING A DA VINCI SI MYOMECTOMY PROCEDURE, THE SURGICAL STAFF REPORTED SEEING FRAYED WIRES ON THE PK DISSECTING FORCEPS INSTRUMENT. NOTHING WAS REPORTED HAVING FALLEN INTO THE PATIENT. NO PATIENT HARM, ADVERSE OUTCOME OR INJURY WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 16410 | PK DISSECTING FORCEPS INSTRUMENT | ENDOSCOPIC ELECTROSURGICAL INSTRUMENT | NAY | INTUITIVE SURGICAL,INC. | 420227-03 | M10121002 916 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | DA VINCI SI SYSTEM INSTRUMENTS AND ACCESSORIES |