FDA Adverse Event Malfunction Summary report: N

PRECISION®

MDR report key: 2912316 · Received January 11, 2013

Report

Report Number
3006630150-2013-00030
Event Type
Malfunction
Date Received
January 11, 2013
Date of Event
December 19, 2012
Report Date
December 19, 2012
Manufacturer
BOSTON SCIENTIFIC NEUROMODULATION
Product Code
LGW
PMA / PMN Number
030017
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE EXPLANTED DEVICES WERE NOT RETURNED TO BSN AS THEY WERE DISCARDED BY THE MEDICAL FACILITY. IT IS INDICATED THAT THE DEVICE WILL NOT BE RETURNED FOR EVALUATION; THEREFORE, A FAILURE ANALYSIS OF THE COMPLAINT DEVICE COULD NOT BE COMPLETED. A REVIEW OF THE DEVICE HISTORY RECORDS WILL BE CONDUCTED. IF THERE IS ANY FURTHER RELEVANT INFORMATION FROM THAT REVIEW, A SUPPLEMENTAL MED WATCH WILL BE FILED.

Description of Event or Problem · 1

A REPORT WAS RECEIVED THAT THE PATIENT UNDERWENT AN IPG REPLACEMENT PROCEDURE DUE TO THE IPG BEING DAMAGED AND NOT PROVIDING STIMULATION. THE PATIENT HAD A NON-DEVICE RELATED BACK SURGERY IN WHICH ELECTROCAUTERY WAS USED. THE PATIENT WAS DOING WELL AFTER THE PROCEDURE. MONOPOLAR ELECTROCAUTERY IS A KNOWN SOURCE OF HIGH VOLTAGE TRANSIENT SIGNAL AND CURRENT COMPANY LABELING WARNS AGAINST THE USE OF MONOPOLAR ELECTROCAUTERY. (PHYSICIAN'S IMPLANT MANUAL 9055940-001).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
17724 PRECISION® SPINAL CORD STIMULATOR LGW BOSTON SCIENTIFIC NEUROMODULATION SC-1110-02 NA

Patients

Seq Age Sex Outcome Treatment
1 53 YR