FDA Adverse Event Malfunction Summary report: N

ENDURANT ABDOMINAL STENT GRAFT SYSTEM

MDR report key: 2912283 · Received January 11, 2013

Report

Report Number
2953200-2013-00074
Event Type
Malfunction
Date Received
January 11, 2013
Date of Event
November 5, 2009
Report Date
January 18, 2013
Manufacturer
MEDTRONIC CARDIOVASCULAR SANTA ROSA
Product Code
MIH
PMA / PMN Number
P100021
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
EZ
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

(B)(4). EVALUATION, RESULTS: INHERENT RISK OF PROCEDURE (ENDOLEAK); PATIENT'S CONDITION AFFECTED EFFECTIVENESS OF DEVICE (CALCIFIED, MILDLY TORTUOUS AND ANEURYSMAL ILIAC ARTERIES). CONCLUSION: DEVICE FAILURE/LACK OF EFFECTIVENESS RELATED TO PATIENT CONDITION (CALCIFIED, MILDLY TORTUOUS AND ANEURYSMAL ILIAC ARTERIES).

Additional Manufacturer Narrative · 1

(B)(4). EVALUATION, RESULTS: INHERENT RISK OF PROCEDURE (ENDOLEAK); PATIENT'S CONDITION AFFECTED EFFECTIVENESS OF DEVICE (CALCIFIED, MILDLY TORTUOUS AND ANEURYSMAL ILIAC ARTERIES). CONCLUSION: DEVICE FAILURE/LACK OF EFFECTIVENESS RELATED TO PATIENT CONDITION (CALCIFIED, MILDLY TORTUOUS AND ANEURYSMAL ILIAC ARTERIES).

Description of Event or Problem · 1

AN ENDURANT STENT GRAFT SYSTEM WAS IMPLANTED IN A PATIENT FOR THE ENDOVASCULAR TREATMENT OF A 67MM DIAMETER X 155MM LENGTH ABDOMINAL AORTIC ANEURYSM APPROXIMATELY 38 MONTHS AGO. THE AORTIC NECK WAS 29MM IN DIAMETER AND 20MM IN LENGTH WITH AN ANGULATION OF 45 DEGREES. THE RIGHT ILIAC ARTERY DIAMETER WAS 19MM AND THE LEFT WAS 25MM. THE LENGTH OF THE RIGHT ILIAC ARTERY ANEURYSM WAS 40MM AND LEFT ILIAC ARTERY ANEURYSM WAS 45MM. THE RIGHT ILIAC ARTERY HAD MILD TORTUOSITY. THERE WAS 40% STENOSIS IN THE ILIAC ARTERIES BILATERALLY. THERE WAS 10% CIRCUMFERENTIAL AORTIC MURAL THROMBUS/CALCIFICATION AT PROXIMAL NECK. IT WAS REPORTED THAT THERE WAS POSITIONING DIFFICULTIES DUE TO A CALCIFIED AND TORTUOUS ILIAC ARTERY; HOWEVER, IT WAS NOT SPECIFIED WHICH ILIAC ARTERY. IT WAS REPORTED THAT THERE WAS ANEURYSM SIZE FLUCTUATION FROM 72MM APPROXIMATELY ONE WEEK POST IMPLANT TO 69MM AT ONE YEAR POST IMPLANT TO 47MM AT THE 2 YEAR FOLLOW-UP TO FINALLY 78MM APPROXIMATELY THREE WEEKS AGO, WHERE THERE WAS A REPORT OF JUNCTIONAL TYPE III ENDOLEAK LEAK/DISCONNECT REPORTED. THERE WAS NO REPORT ON WHICH STENT GRAFTS SEPARATED AND NO INTERVENTION HAS BEEN PERFORMED. THE PATIENT IS BEING MONITORED.

Description of Event or Problem · 1

IT WAS REPORTED TYPE III ENDOLEAK IS BETWEEN THE BIFURCATED STENT GRAFT/MAIN BODY AND THE CONTRALATERAL LIMB IN RIGHT ILIAC ARTERY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
17932 ENDURANT ABDOMINAL STENT GRAFT SYSTEM SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT MIH MEDTRONIC CARDIOVASCULAR SANTA ROSA V00289707

Patients

Seq Age Sex Outcome Treatment
1 00067 YR