FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME INSULIN INFUSION PUMP

MDR report key: 2912253 · Received January 11, 2013

Report

Report Number
3004209178-2013-90232
Event Type
Malfunction
Date Received
January 11, 2013
Date of Event
December 27, 2012
Report Date
December 27, 2012
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MN
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT. THE DEVICE HAS BEEN RETURNED, BUT NOT YET EVALUATED. NO CONCLUSION CAN BE DRAWN AT THIS TIME. FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED.

Additional Manufacturer Narrative · 1

THE INSULIN PUMP PASSED THE REWIND AND DISPLACEMENT TEST. HOWEVER, UNABLE TO PRIME THE INSULIN PUMP DURING PRIME TEST DUE TO FAULTY FORCE SENSOR. UNABLE TO PERFORM BASIC OCCLUSION TEST, OCCLUSION, EXCESSIVE NO DELIVERY TEST OR VERIFY MOTOR ERROR ALARMS DUE TO PRIME ANOMALY. THE MOTOR WAS TESTED OUTSIDE THE DEVICE AND PASSED.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT THE INSULIN PUMP ALARMED MOTOR ERROR. THE BLOOD GLUCOSE READING WAS UNKNOWN. TROUBLESHOOTING WAS PERFORMED. THE CALLER STATED THAT THE DEVICE WAS NOT EXPOSED TO A HIGH MAGNETIC FIELD. ASSISTED THE CUSTOMER TO RUN THE DISPLACEMENT TEST AND THE TEST FAILED. ADVISED THE CALLER THAT THE INSULIN PUMP WOULD BE REPLACED AND REVERT TO BACK UP PLAN. NO FURTHER INFORMATION WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
16569 PARADIGM REAL-TIME INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MMT-522NAL

Patients

Seq Age Sex Outcome Treatment
1 50 YR