AU680 CHEMISTRY ANALYZER WITH ISE
Report
- Report Number
- 9612296-2013-00004
- Event Type
- Malfunction
- Date Received
- January 11, 2013
- Date of Event
- December 17, 2012
- Report Date
- December 17, 2012
- Manufacturer
- BECKMAN COULTER INC.
- Product Code
- JJE
- PMA / PMN Number
- K961274
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NY, US
- Reporter Occupation
- OTHER
Narratives
THE CUSTOMER REPORTED THAT THE ORIGINAL TP CALIBRATION AND QC WERE ALL FOUND TO BE ACCEPTABLE BEFORE THE EVENT. BECKMAN COULTER FIELD SERVICE ENGINEER (FSE) WAS DISPATCHED TO THE CUSTOMER SITE. THE FSE DISCOVERED THE TP PRECISION ISSUES. THE FSE REPLACED THE R2 SYRINGE AND PROBE AND THEN RAN WATER BLANK TEST WHICH PASSED. THE TP ASSAY WAS RECALIBRATED AND QC WAS PERFORMED; ALL RESULTS PASSED. TO DATE, THERE HAVE BEEN NO FURTHER TP ISSUES REPORTED FROM THIS CUSTOMER SITE. R2 REAGENT PROBE AND SYRINGE REPLACEMENT FIXED THE PROBLEM.
A CUSTOMER CONTACTED BECKMAN COULTER INC., (BCI) AND REPORTED OBTAINING LOW TOTAL PROTEIN (TP) RESULTS FOR PATIENT SAMPLES WHICH WERE GENERATED FROM AN AU680 CHEMISTRY ANALYZER. THE CUSTOMER STATED THAT THEY WERE QUESTIONED BY SEVERAL DOCTORS ABOUT TP RESULTS THAT WERE LOWER THAN EXPECTED. CUSTOMER NOTED THAT THE LOW TP RESULTS WERE SHOWING UP AFTER HAVING PROCESSED ABOUT 150 PATIENT SAMPLES. THE CUSTOMER PROVIDED ONLY THE FOLLOWING EXAMPLES OF PATIENT RESULTS: 5.9, 5.7, 5.4 WITH A RERUN OF 6.0, 6.7 6.5 RESPECTIVELY. THE CUSTOMER DID NOT PROVIDE ANY PATIENT RESULTS DOCUMENTATION FOR THIS INVESTIGATION. THE CUSTOMER DID NOT INDICATE THE NUMBER OF PATIENT RESULTS THAT WERE ERRONEOUS. ONCE THE LOW TP RESULTS WERE DISCOVERED, ALL SAMPLES WERE REPEATED BACK TO THE PREVIOUS QC RUN. REPEATED RESULTS WERE PERFORMED ON THEIR OTHER INSTRUMENT. CUSTOMER STATED THAT SOME AMENDED REPORTS HAD BEEN ISSUED. PATIENT RESULTS WERE RELEASED OUT OF THE LAB. ALTHOUGH THE CUSTOMER DID NOT RECEIVE REPORTS OF CHANGE TO PATIENT TREATMENT, IT IS UNKNOWN IF ANY PATIENT WAS AFFECTED IN CONNECTION WITH THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 18038 | AU680 CHEMISTRY ANALYZER WITH ISE | ANALYZER, CHEMISTRY (PHOTOMETRIC, DISCRETE), FOR CLINICAL USE | JJE | BECKMAN COULTER INC. | AU681-02E | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |