FDA Adverse Event Malfunction Summary report: N

FLOGARD

MDR report key: 2912124 · Received January 11, 2013

Report

Report Number
1416980-2013-00961
Event Type
Malfunction
Date Received
January 11, 2013
Date of Event
December 24, 2012
Report Date
December 24, 2012
Manufacturer
BAXTER HEALTHCARE - SINGAPORE
Product Code
FRN
PMA / PMN Number
K915522
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PR
Reporter Occupation
SERVICE AND TESTING PERSONNEL

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE DEVICE HAS BEEN RECEIVED BY BAXTER, AND THE CONDITION WAS CONFIRMED. THIS IS AN ANCILLARY SERVICE EVENT. THE CAUSE WAS DETERMINED TO BE A FAULTY SAFETY SLIDE CLAMP ASSEMBLY. THE DEVICE WAS RETURNED UNREPAIRED DUE TO PART OBSOLESCENCE. IF ANY ADDITIONAL INFORMATION IS RECEIVED, A FOLLOW UP MDR WILL BE SENT.

Description of Event or Problem · 1

DURING PRODUCT EVALUATION BY A BAXTER SERVICE TECHNICIAN, A FLOGARD VOLUMETRIC INFUSION PUMP WAS FOUND TO HAVE AN ALARM F27. THIS WAS NOT REPORTED BY THE CUSTOMER. THIS CONDITION HAD THE POTENTIAL TO INTERRUPT DELIVERY. THERE WAS NO PATIENT INVOLVEMENT; THEREFORE, NO PATIENT INJURY, MEDICAL INTERVENTION, OR ADVERSE REACTION IS ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
16362 FLOGARD PUMP, INFUSION FRN BAXTER HEALTHCARE - SINGAPORE

Patients

Seq Age Sex Outcome Treatment
1