FDA Adverse Event Injury Summary report: N

ENDURANT ABDOMINAL STENT GRAFT SYSTEM

MDR report key: 2912112 · Received January 11, 2013

Report

Report Number
2953200-2013-00066
Event Type
Injury
Date Received
January 11, 2013
Date of Event
December 11, 2012
Report Date
January 14, 2013
Manufacturer
MEDTRONIC IRELAND
Product Code
MIH
PMA / PMN Number
P100021
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
FR
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

(B)(4). EVALUATION, RESULTS: INHERENT RISK OF PROCEDURE (STENT GRAFT OCCLUSION), PATIENT'S CONDITION AFFECTED EFFECTIVENESS OF DEVICE (TAPERED SHAPED AORTIC NECK). EVALUATION, CONCLUSION: DEVICE FAILURE/LACK OF EFFECTIVENESS RELATED TO PATIENT CONDITION (TAPERED SHAPED AORTIC NECK).

Description of Event or Problem · 1

AN ENDURANT STENT GRAFT SYSTEM WAS IMPLANTED IN A PATIENT FOR THE ENDOVASCULAR TREATMENT OF A 51 MM DIAMETER ABDOMINAL AORTIC ANEURYSM AND 17 MM DIAMETER ILIAC ARTERY ANEURYSM APPROXIMATELY EIGHT MONTHS AGO. THE AORTIC NECK WAS TAPERED SHAPED MEASURING 22-27MM AND FROM PROXIMAL TO DISTAL AND 25 MM IN LENGTH. DIAMETER OF THE ILIAC ARTERIES WAS 14 MM BILATERALLY. THE FEMORAL ARTERIES WERE 8 MM IN DIAMETER BILATERALLY. THERE WAS NO PRESENCE OF CIRCUMFERENTIAL THROMBUS OR CALCIFICATION. POST-OP CT IMAGING SHOWED NO ADVERSE EVENTS AND ANEURYSM SIZE WAS 51MM. THE PATIENT WAS DISCHARGED FOUR DAYS POST IMPLANT. A RECENT FOLLOW-UP CT SCAN FROM APPROXIMATELY ONE MONTH AGO SHOWED THE ANEURYSM SIZE WAS 44MM AND CIRCUMFERENTIAL THROMBUS GROWTH AT THE LEVEL OF THE DEVICE WAS REPORTED AND NOTED AS WAS PRESENT AT PREVIOUS FILM READ ON A PREVIOUS UNKNOWN DATE. NO INTERVENTION PERFORMED AND THE PATIENT TO BE MONITORED.

Description of Event or Problem · 1

ADDITIONAL INFORMATION WAS RECEVIED AS FOLLOWS: PER THE INVESTIGATOR THERE WAS CIRCUMFERENTIAL THROMBUS WITHIN THE STENT GRAFT OF THE MAIN BODY. THIS MDR IS REDACTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
17068 ENDURANT ABDOMINAL STENT GRAFT SYSTEM SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT MIH MEDTRONIC IRELAND V01043501

Patients

Seq Age Sex Outcome Treatment
1 00061 YR