FDA Adverse Event Malfunction Summary report: N

FLO-THRU INTRALUMINAL SHUNT

MDR report key: 2912099 · Received January 11, 2013

Report

Report Number
2032282-2013-00002
Event Type
Malfunction
Date Received
January 11, 2013
Date of Event
December 25, 2012
Report Date
December 25, 2012
Manufacturer
SYNOVIS SURGICAL INNOVATIONS
Product Code
DXC
PMA / PMN Number
K981624
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

(B)(4). VISUAL INSPECTION OF THE COMPLAINT SAMPLE BY BAXTER SYNOVIS OBSERVED A SINGLE BLACK FIBER BETWEEN THE INNER AND OUTER POUCHES (NO DIRECT CONTACT WITH DEVICE) APPROXIMATELY 4.1MM X 0.9MM IN SIZE. PARTICULATE ANALYSIS BY BAXTER ROUND LAKE IDENTIFIED THE PARTICLE AS A PIECE OF STEEL FLASHING. THE COMPLAINT WAS CONFIRMED. BATCH RECORD REVIEW BY BAXTER SYNOVIS FOUND NO ISSUES THAT COULD HAVE CONTRIBUTED TO THIS COMPLAINT ISSUE. ALL PRODUCT REQUIREMENTS WERE MET AND THE LOT PASSED ALL INSPECTIONS. THERE WERE NO ISSUES AND NO SCRAP REPORTED. PER BAXTER SYNOVIS, THE PRODUCT IS INSPECTED FOR FOREIGN MATERIAL AT THREE POINTS DURING MANUFACTURE. NO TREND WAS IDENTIFIED. BAXTER SYNOVIS DETERMINED NO MANUFACTURING DEFECT WAS FOUND, THE ROOT CAUSE IS INDETERMINABLE, AND THE ISSUE IS CONSIDERED LOW RISK. PER ADDITIONAL INFORMATION RECEIVED FROM BAXTER SYNOVIS, (B)(4) HAS BEEN INITIATED TO FURTHER INVESTIGATE PARTICULATE MATTER ISSUES. THIS IS THE FIRST REPORTED COMPLAINT OF THIS NATURE FOR THIS PRODUCT LOT. THIS COMPLAINT WILL BE KEPT ON FILE FOR TRENDING PURPOSES.

Additional Manufacturer Narrative · 1

(B)(4). BAXTER MEDICAL ASSESSMENT: THE FOREIGN BODY IN THE PRODUCT POUCH WAS OBVIOUSLY IDENTIFIED PRIOR TO USE, SO IN THIS CASE THERE WAS NO POSSIBLE HARM OR INJURY OF PATIENT. IF SUCH TYPE OF MALFUNCTION OR HAZARD WOULD RECUR AND USER DID NOT RECOGNIZE THE FOREIGN BODY PRIOR TO USE, IT POSSIBLY CAN BE TRANSFERRED TO THE TREATMENT SITE OF THE PATIENT. IN A WORST CASE SCENARIO, THIS MAY LEAD TO A FOREIGN BODY REACTION AND STERILE VASCULAR INFLAMMATION OR THROMBOTIC REACTION WHICH ONLY IN EXCEPTIONAL CASES MAY CAUSE SERIOUS HARM OR INJURY. ADDITIONAL INVESTIGATION IS CURRENTLY ONGOING AND THE SAMPLE HAS BEEN REQUESTED FOR EVALUATION. UPON ITS COMPLETION A FOLLOW-UP SUBMISSION WILL BE SENT.

Description of Event or Problem · 1

CUSTOMER REPORTED: WE HAVE FOUND FOREIGN MATTER IN POUCHES OF FLO-THRU. NO ADDITIONAL INFORMATION PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
17799 FLO-THRU INTRALUMINAL SHUNT CLAMP, VASCULAR DXC SYNOVIS SURGICAL INNOVATIONS 5802023-1838080

Patients

Seq Age Sex Outcome Treatment
1