ENDOTAK RELIANCE
Report
- Report Number
- 2124215-2012-16537
- Event Type
- Injury
- Date Received
- January 11, 2013
- Date of Event
- October 4, 2012
- Report Date
- December 18, 2012
- Manufacturer
- CPI - DEL CARIBE
- Product Code
- NVY
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IL
- Reporter Occupation
- PHYSICIAN
Narratives
THE PRODUCT HAS BEEN RECEIVED FOR ANALYSIS. THIS REPORT WILL BE UPDATED UPON COMPLETION OF ANALYSIS.
UPON RECEIPT AT OUR POST MARKET QUALITY ASSURANCE LABORATORY, A THOROUGH EVALUATION OF THE LEAD WAS PERFORMED. VISUAL INSPECTION NOTED TRILUMEN INSULATION ABRASION EXPOSING THE RATE SENSE NEGATIVE COIL APPROXIMATELY 49-51CM FROM THE TERMINAL PIN. UNDER X-RAY, THE ABRASION DAMAGE APPEARED TO BE FROM LEAD ON VALVE CONTACT. THE HELIX MECHANISM RETRACTION DIFFICULTY WAS LIKELY DUE TO BLOOD INFILTRATION INTO THE MECHANISM BECAUSE OF THE BREACH IN INSULATION. THE LEAD REMAINED ELECTRICALLY CONTINUOUS.
BOSTON SCIENTIFIC RECEIVED INFORMATION THAT THIS LEAD EXHIBITED LOW OUT OF RANGE PACING IMPEDANCE MEASUREMENTS (LESS THAN 200 OHMS). DURING LEAD EVALUATION, SENSING AND THRESHOLD MEASUREMENTS WERE NOTED TO BE STABLE AND WITHIN NORMAL LIMITS. THE DECISION WAS MADE LEAVE THE LEAD IN SERVICE AND MONITOR. TWO MONTHS LATER, IT WAS REPORTED THAT THIS TRANSVENOUS PACING SYSTEM WOULD BE REPLACED WITH AN S-ICD SYSTEM (BECAUSE OF THE LOW IMPEDANCE MEASUREMENTS). DURING THE REVISION PROCEDURE, THE CHRONIC RIGHT VENTRICULAR (RV) LEAD WAS TESTED THROUGH THE PACING SYSTEM ANALYZER (PSA) AND STILL YIELDED LOW PACING IMPEDANCE MEASUREMENTS. THE LEAD WAS THEN EXTRACTED. DURING THE EXTRACTION, DIFFICULTY WAS ENCOUNTERED REMOVING THE LEAD TIP. THE EZ4 CONNECTOR TOOL WAS USED HOWEVER THE HELIX WOULD NOT RETRACT WITH 20 TURNS. THE PHYSICIAN THEN TORQUED THE ENTIRE LEAD WHICH FREED THE LEAD TIP ENOUGH TO THEN USE THE EZ4 CONNECTOR TO RELEASE THE HELIX AND REMOVE THE LEAD. UPON THE LEAD BEING REMOVED, IT WAS NOTED A SIGNIFICANT ABRASION JUST PROXIMAL TO THE COIL (SINGLE COIL LEAD) WAS SEEN. THE PATIENT DOES HAVE A MECHANICAL MITRAL VALVE HOWEVER HAS HAD NO PROCEDURES ON HER TRICUSPID VALVE. X-RAY SHOWED THAT THIS PORTION OF THE LEAD WAS LIKELY LAYING ACROSS THE TRICUSPID VALVE. THE ABRASION APPEARS LIKE IT COULD HAVE BEEN A CUT WITH A SCALPEL BUT NO KNOWN ISSUES WERE REPORTED AT THE IMPLANT AND IMPEDANCE MEASUREMENTS WERE REPORTED TO BE WITHIN NORMAL LIMITS DURING IMPLANT TESTING (HOWEVER THE LEAD WAS IMPLANTED WITH A COMPETITOR DEVICE AND NO BOSTON SCIENTIFIC FIELD REPRESENTATIVE WAS PRESENT).
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 16670 | ENDOTAK RELIANCE | NVY | CPI - DEL CARIBE | 0292 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 51 YR | Hospitalization| L| R | MISMATCH| 0292 |