FDA Adverse Event Malfunction Summary report: N

VITALITY 2

MDR report key: 2911936 · Received January 11, 2013

Report

Report Number
2124215-2012-16327
Event Type
Malfunction
Date Received
January 11, 2013
Date of Event
December 6, 2012
Report Date
December 7, 2012
Manufacturer
GUIDANT CRM CLONMEL IRELAND
Product Code
LWS
Removal / Correction Number
TR06026
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

UPON COMPLETION OF ANALYSIS, THIS REPORT WILL BE UPDATED.

Additional Manufacturer Narrative · 1

UPON RECEIPT AT OUR POST MARKET QUALITY ASSURANCE LABORATORY, A REVIEW OF DEVICE MEMORY REVEALED THAT THE DEVICE DECLARED ELECTIVE REPLACEMENT INDICATOR (ERI) AFTER EXPERIENCING TWO CONSECUTIVE CHARGE TIMES GREATER THAN THE EXTENDED MID-LIFE ERI CHARGE TIME LIMIT. THE OBSERVED RATE OF BATTERY USAGE WAS COMPARED TO THE EXPECTED RATE OF BATTERY USAGE; THE RESULTS INDICATED THAT THE MONITORING VOLTAGE WAS NORMAL. ALTHOUGH THE BATTERY ITSELF HAD NOT DEPLETED PREMATURELY, ERI WAS TRIGGERED EARLIER THAN EXPECTED BY EXTENDED CHARGE TIMES DUE TO A HIGHER-THAN-TYPICAL BUILDUP OF INTERNAL BATTERY IMPEDANCE.

Description of Event or Problem · 1

BOSTON SCIENTIFIC CRM RECEIVED INFORMATION THAT THIS IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD) WAS RETURNED FOR DISPOSAL AFTER EXPLANT. NO PRODUCT PERFORMANCE ALLEGATIONS OR ADVERSE PATIENT EFFECTS HAVE BEEN REPORTED WITH REGARD TO THIS DEVICE. ROUTINE INITIAL RETURNED DEVICE TESTING INDICATED THAT DETAILED ANALYSIS WAS REQUIRED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
18322 VITALITY 2 IMPLANTABLE CARDIOVERTER DEFIBRILLATOR LWS GUIDANT CRM CLONMEL IRELAND T177

Patients

Seq Age Sex Outcome Treatment
1 51 YR 0185| T177| E141