REJUVENATE STRGHT PRFIT TMZF MOD STEM SIZE 7
Report
- Report Number
- 0002249697-2013-00027
- Event Type
- Injury
- Date Received
- January 11, 2013
- Date of Event
- February 23, 2012
- Report Date
- November 14, 2012
- Manufacturer
- STRYKER ORTHOPAEDICS-MAHWAH
- Product Code
- MEH
- PMA / PMN Number
- K092561
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FL, US
- Reporter Occupation
- PHYSICIAN
Narratives
AN EVALUATION OF THE DEVICE CANNOT BE PERFORMED AS THE DEVICE WAS NOT RETURNED TO THE MANUFACTURER. ADDITIONAL INFORMATION (INCLUDING X-RAYS AND MEDICAL RECORDS) HAS BEEN REQUESTED. SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE, THE EVALUATION SUMMARY WILL BE SUBMITTED IN A SUPPLEMENTAL REPORT. DEVICE EVALUATED BY MFR: NOT RETURNED TO THE MANUFACTURER.
AN EVENT REGARDING REVISION DUE TO ELEVATED COBALT/CHROME INVOLVING AN UNKNOWN REJUVENATE MODULAR DEVICE WAS REPORTED. THE EVENT WAS CONFIRMED. DEVICE EVALUATION WAS NOT PERFORMED AS NO DEVICES WERE RECEIVED. REVIEW OF DEVICE HISTORY RECORDS COULD NOT BE PERFORMED AS THE REPORTED DEVICE WAS NOT PROPERLY IDENTIFIED. SIMILAR EVENTS HAVE OCCURRED FOR THE REJUVENATE MODULAR PRODUCT FAMILY. THESE EVENTS WERE DETERMINED TO BE ASSOCIATED WITH RA 2012-067. VOLUNTARY RECALL RA 2012-067 WAS INITIATED FOR ABGII AND REJUVENATE MODULAR STEMS AND NECKS DUE TO THE POTENTIAL RISKS ASSOCIATED WITH THESE DEVICES. THE REPORTED REVISION DUE TO ELEVATED COBALT/CHROME IS CONSIDERED TO BE UNDER THE SCOPE OF THIS RECALL
DR (B)(6) REPORTS PATIENT IS EXPERIENCING SYMPTOMS RELATED TO ELEVATED COBALT/CHROME, INCLUDING PSEUDOTUMOR. POSITIVE MARS SCAN.
DR. (B)(6) REPORTS PATIENT IS EXPERIENCING SYMPTOMS RELATED TO ELEVATED COBALT/CHROME, INCLUDING PSEUDOTUMOR. POSITIVE MARS SCAN.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 18467 | REJUVENATE STRGHT PRFIT TMZF MOD STEM SIZE 7 | IMPLANT | MEH | STRYKER ORTHOPAEDICS-MAHWAH | UNK |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 79 YR | Required Intervention |