FDA Adverse Event Injury Summary report: N

LRG TAP PRI MOD NCK 8DEG 30MM

MDR report key: 2911713 · Received January 11, 2013

Report

Report Number
0002249697-2013-00019
Event Type
Injury
Date Received
January 11, 2013
Date of Event
August 23, 2011
Report Date
August 22, 2012
Manufacturer
STRYKER ORTHOPAEDICS-MAHWAH
Product Code
MEH
PMA / PMN Number
K092561
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

AN EVALUATION OF THE DEVICE CANNOT BE PERFORMED AS THE DEVICE WAS NOT RETURNED TO THE MANUFACTURER. ADDITIONAL INFORMATION (INCLUDING X-RAYS AND MEDICAL RECORDS) HAS BEEN REQUESTED. SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE, THE EVALUATION SUMMARY WILL BE SUBMITTED IN A SUPPLEMENTAL REPORT.

Additional Manufacturer Narrative · 1

AN EVENT REGARDING ALTR INVOLVING A REJUVENATE MODULAR DEVICE WAS REPORTED. THE EVENT WAS CONFIRMED. DEVICE EVALUATION WAS NOT PERFORMED AS NO DEVICES WERE RECEIVED. A DEVICE HISTORY REVIEW FOUND THAT THE DEVICES IN THE REPORTED LOT WERE ACCEPTED INTO FINAL STOCK WITH NO REPORTED DISCREPANCIES. THE SUPER AND CHS COMPLAINT DATABASES SHOW THERE HAVE BEEN OTHER EVENTS FOR THE REPORTED LOT. SIMILAR EVENTS HAVE OCCURRED FOR THE CATALOG NUMBER AND PRODUCT FAMILY. THESE EVENTS WERE DETERMINED TO BE ASSOCIATED WITH RA 2012-067. VOLUNTARY RECALL RA 2012-067 WAS INITIATED FOR ABGII AND REJUVENATE MODULAR STEMS AND NECKS DUE TO THE POTENTIAL RISKS ASSOCIATED WITH THESE DEVICES. THE REPORTED ALTR IS CONSIDERED TO BE UNDER THE SCOPE OF THIS RECALL. NO FURTHER INVESTIGATION IS REQUIRED.

Description of Event or Problem · 1

DR. (B)(6) REPORTS PATIENT IS EXPERIENCING SYMPTOMS RELATED TO ELEVATED COBALT/CHROME, INCLUDING PSEUDOTUMOR. POSITIVE MARS SCAN.

Description of Event or Problem · 1

DR. (B)(6) REPORTS PATIENT IS EXPERIENCING SYMPTOMS RELATED TO ELEVATED COBALT/CHROME, INCLUDING PSEUDOTUMOR. POSITIVE MARS SCAN.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
17441 LRG TAP PRI MOD NCK 8DEG 30MM IMPLANT MEH STRYKER ORTHOPAEDICS-MAHWAH 36168601

Patients

Seq Age Sex Outcome Treatment
1 79 YR Required Intervention