FDA Adverse Event Malfunction Summary report: N

CORONARY SINUS CATHETER

MDR report key: 2911689 · Received January 11, 2013

Report

Report Number
3008500478-2013-00363
Event Type
Malfunction
Date Received
January 11, 2013
Date of Event
December 28, 2012
Report Date
February 11, 2013
Manufacturer
EDWARDS LIFESCIENCES
Product Code
DWF
PMA / PMN Number
5120780
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

SINCE THE DEVICE WAS NOT RETURNED, ANY MANUFACTURING DEFECT WITH THE DEVICE CANNOT BE ASSESSED. HOWEVER, IT IS LIKELY THAT THE ROOT CAUSE OF THE INTRODUCER LEAKING IS RELATED TO THE ROOT CAUSE ATTRIBUTED TO SIMILAR COMPLAINTS RECEIVED FOR THE LEAKING THAT HAVE BEEN CONFIRMED PER ASSESSMENT BY R&D AND QUALITY ENGINEERING OF THE SIMILAR COMPLAINTS WITH RETURNED DEVICES, THE ROOT CAUSE OF THE LEAKING HAS BEEN DETERMINED TO BE A MATERIAL RELAXATION ISSUE. IT IS NOT KNOWN AT THIS TIME IF THE VALVE LEAKING IS A DESIGN OR A SUPPLIER MANUFACTURING ISSUE. A CAPA HAS BEEN INITIATED FOR THESE COMPLAINTS AND PRODUCT RECALL. TRENDS WILL CONTINUE TO BE MONITORED ON A MONTHLY BASIS AND IF FURTHER ACTION IS REQUIRED, APPROPRIATE INVESTIGATION WILL BE PERFORMED.

Additional Manufacturer Narrative · 1

THE PRODUCT WAS NOT RETURNED AND THE ROOT CAUSE COULD NOT BE DETERMINED FOR THIS DEVICE. THEREFORE, A CAPA WAS NOT INITIATED FOR THIS EVENT. THIS INFORMATION WILL BE INCLUDED IN TRENDING AND FUTURE CAPA DETERMINATIONS.

Description of Event or Problem · 1

IT WAS PHONED IN BY THE SALES REP THAT THE MD EXPERIENCED SIGNIFICANT LEAKAGE OUT OF THE INTRODUCER AFTER HE INSERTED THE CORONARY SINUS CATHETER, PR9. HE TRIED REINSERTING THE DEVICE BUT IT STILL LEAKED AND HAD TO MANAGE THE LEAKING DURING THE CASE. THEY DID NOT SAVE THE NON FUNCTIONING ONE. NO PATIENT ISSUES OCCURRED. THEY DID OPEN A SECOND INTRODUCER FOR THE CASE. NO PATIENT INJURY WAS REPORTED

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
18610 CORONARY SINUS CATHETER CATHETER, CANNULA AND TUBING, VASCULAR, CARDIOPULMONARY BYPASS DWF EDWARDS LIFESCIENCES PR9

Patients

Seq Age Sex Outcome Treatment
1