CD 1800 ANALYZER
Report
- Report Number
- 2919069-2008-00480
- Event Type
- Malfunction
- Date Received
- March 26, 2008
- Date of Event
- March 4, 2008
- Report Date
- March 5, 2008
- Manufacturer
- ABBOTT DIAGNOSTICS DIVISION/CELLDYN
- Product Code
- GKZ
- PMA / PMN Number
- K030513
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GA, US
- Reporter Occupation
- OTHER
Narratives
(B)(4). RESULT, OTHER: SILICON TUBING (S1 - BUBBLE MIX, S2 - PINCH TUBING, S4 TUBING), RETAINER FOR PROBE, WASTE OUTLET TUBE ASSEMBLY (TUBING AND SENSOR WIRE). THE FIELD SERVICE REP (FSR) VISITED ON (B)(4) 2008 AND OBSERVED ISSUES WITH FLOW/CLOGS. SEVEN PARTS WERE REPLACED. THE S1 (SILICON) TUBING FOR THE RBC/PLT BUBBLE MIX WAS REPLACED AS CLOGGED/OBSTRUCTED (HARD FAILURE). S2 SILICON TUBING FOR THE FLOW PANEL PINCH TUBING WAS REPLACED AS WORN OUT (PREVENTIVE MAINTENANCE). S4 SILICON TUBING WAS CHANGED AS A PROACTIVE REPLACEMENT (PREVENTIVE MAINTENANCE). WORN OUT SYRINGES WERE REPLACED (100 UL SAMPLE SYRINGE-PREVENTIVE MAINTENANCE, LEAKING 10 ML DILUENT SYRINGE-HARD FAILURE) THE WASTE CABLE/TUBING WAS REPLACED AS ONE WIRE WAS BROKEN (HARD FAILURE). THE RETAINER FOR THE ASPIRATION PROBE WAS MISSING AND ALSO REPLACED (HARD FAILURE). PRECISION AND CONTROLS WERE RUN AND THE INSTRUMENT WAS PERFORMING WITHIN LABELING CLAIMS. THE CELL-DYN SYSTEM 1800 OPERATOR'S MANUAL, 07H80-01, REV D, PROVIDES A TROUBLESHOOTING GUIDE IN THE INTRODUCTION. A REVIEW OF THE COMPLAINT ANALYSIS TRENDING SYSTEM (CATS) AND TRENDING ANALYSIS SUPPORT SYSTEM (TASS) REPORTS, FOR THE PERIOD JUNE 2007 THROUGH JANUARY 2008, DID NOT INDICATE ANY ADVERSE TREND WITH THE CELL-DYN 1800, LIST BASE 07H77-01, FOR ISSUES WITH DISCREPANT RESULTS ON MULTIPLE PARAMETERS. THERE WAS NO SYSTEMIC ISSUE. THS IS A FINAL REPORT. END OF REPORT.
THE ACCOUNT STATED THAT ONE PT SPECIMEN TESTED CELL-DYN 1800 HEMOGLOBIN = 8.3 G/DL BUT TESTED HEMOGLOBIN = 10.8 G/DL AT A REFERENCE LABORATORY. THE ACCOUNT SUSPECTED AN ASPIRATION ISSUE ON THE CELL-DYN 1800, AS THE SAMPLE HAD TO BE HELD UNDER THE PROBE FOR AN EXTENDED PERIOD BEFORE SAMPLE ASPIRATION BEGAN. ADDITIONALLY, THE ACCOUNT NOTICED SALT BUILD UP ON THE DILUENT SYRINGE. THE CONTROLS WERE RUNNING WITHIN RANGE AND THE MAINTENANCE WAS UP TO DATE ON THE CELL-DYN 1800. SERVICE WAS REQUESTED FOR THE CELL-DYN 1800. THE ACCOUNT DID NOT REPORT THE CELL-DYN 1800 HEMOGLOBIN RESULT OUTSIDE OF THE LABORATORY. NO IMPACT TO PT MANAGEMENT WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | CD 1800 ANALYZER | AUTOMATED HEMATOLOGY ANALYZER | GKZ | ABBOTT DIAGNOSTICS DIVISION/CELLDYN | NA | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNK |