FINELINE II
Report
- Report Number
- 2124215-2012-16354
- Event Type
- Injury
- Date Received
- January 11, 2013
- Date of Event
- November 21, 2012
- Report Date
- February 20, 2013
- Manufacturer
- CPI - DEL CARIBE
- Product Code
- NVN
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- MEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE
Narratives
AT THIS TIME, THIS LEAD HAS NOT BEEN RETURNED. IF ADDITIONAL INFORMATION IS RECEIVED, THIS REPORT WILL BE UPDATED.
UPON RECEIPT AT OUR POST MARKET QUALITY ASSURANCE LABORATORY, A THOROUGH EVALUATION OF THE LEAD WAS PERFORMED. RESISTANCE AND PRESSURE TESTS WERE COMPLETED TO ASSESS LEAD ELECTRICAL PERFORMANCE AND INSULATION INTEGRITY. MEASUREMENTS THROUGHOUT THESE TESTS WERE WITHIN NORMAL LIMITS. MICROSCOPIC INSPECTIONS OF THE TERMINAL PIN ASSEMBLY, LEAD BODY, AND ELECTRODE TIP FOUND NO ANOMALIES. LABORATORY TESTING WAS UNABLE TO REPRODUCE THE REPORTED CLINICAL OBSERVATIONS AND DETAILED ANALYSIS DID NOT REVEAL ANY ABNORMALITIES.
BOSTON SCIENTIFIC RECEIVED INFORMATION THAT THIS RIGHT VENTRICULAR (RV) LEAD LEAD BECAME DISLODGED. A REVISION PROCEDURE WAS PERFORMED AND THE LEAD WAS EXPLANTED. NO FURTHER ADVERSE PATIENT EFFECTS WERE REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 18839 | FINELINE II | IMPLANTABLE LEAD | NVN | CPI - DEL CARIBE | 4457 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 82 YR | Hospitalization| L| R | K173| 4136| 4457 |