FDA Adverse Event Injury Summary report: N

FINELINE II

MDR report key: 2911550 · Received January 11, 2013

Report

Report Number
2124215-2012-16354
Event Type
Injury
Date Received
January 11, 2013
Date of Event
November 21, 2012
Report Date
February 20, 2013
Manufacturer
CPI - DEL CARIBE
Product Code
NVN
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
MEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE

Narratives

Additional Manufacturer Narrative · 1

AT THIS TIME, THIS LEAD HAS NOT BEEN RETURNED. IF ADDITIONAL INFORMATION IS RECEIVED, THIS REPORT WILL BE UPDATED.

Additional Manufacturer Narrative · 1

UPON RECEIPT AT OUR POST MARKET QUALITY ASSURANCE LABORATORY, A THOROUGH EVALUATION OF THE LEAD WAS PERFORMED. RESISTANCE AND PRESSURE TESTS WERE COMPLETED TO ASSESS LEAD ELECTRICAL PERFORMANCE AND INSULATION INTEGRITY. MEASUREMENTS THROUGHOUT THESE TESTS WERE WITHIN NORMAL LIMITS. MICROSCOPIC INSPECTIONS OF THE TERMINAL PIN ASSEMBLY, LEAD BODY, AND ELECTRODE TIP FOUND NO ANOMALIES. LABORATORY TESTING WAS UNABLE TO REPRODUCE THE REPORTED CLINICAL OBSERVATIONS AND DETAILED ANALYSIS DID NOT REVEAL ANY ABNORMALITIES.

Description of Event or Problem · 1

BOSTON SCIENTIFIC RECEIVED INFORMATION THAT THIS RIGHT VENTRICULAR (RV) LEAD LEAD BECAME DISLODGED. A REVISION PROCEDURE WAS PERFORMED AND THE LEAD WAS EXPLANTED. NO FURTHER ADVERSE PATIENT EFFECTS WERE REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
18839 FINELINE II IMPLANTABLE LEAD NVN CPI - DEL CARIBE 4457

Patients

Seq Age Sex Outcome Treatment
1 82 YR Hospitalization| L| R K173| 4136| 4457