FDA Adverse Event Injury Summary report: N

EASYTRAK 2

MDR report key: 2911437 · Received January 11, 2013

Report

Report Number
2124215-2012-16306
Event Type
Injury
Date Received
January 11, 2013
Date of Event
November 27, 2012
Report Date
December 20, 2012
Manufacturer
CPI - DEL CARIBE
Product Code
OJX
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
UT
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

ALL AVAILABLE INFORMATION INDICATES THAT THIS PRODUCT WAS EXPLANTED. THIS INVESTIGATION WILL BE UPDATED SHOULD FURTHER INFORMATION BE PROVIDED.

Additional Manufacturer Narrative · 1

UPON RECEIPT AT OUR POST MARKET QUALITY ASSURANCE LABORATORY, A THOROUGH EVALUATION OF THE LEAD WAS PERFORMED. RESISTANCE AND PRESSURE TESTS WERE COMPLETED TO ASSESS LEAD ELECTRICAL PERFORMANCE AND INSULATION INTEGRITY. MEASUREMENTS THROUGHOUT THESE TESTS WERE WITHIN NORMAL LIMITS. MICROSCOPIC INSPECTIONS OF THE TERMINAL PIN ASSEMBLY, LEAD BODY, AND ELECTRODE TIP FOUND NO ANOMALIES. LABORATORY TESTING WAS UNABLE TO REPRODUCE THE REPORTED CLINICAL OBSERVATIONS AND DETAILED ANALYSIS DID NOT REVEAL ANY ABNORMALITIES. LABORATORY ANALYSIS DID NOT IDENTIFY ANY LEAD CHARACTERISTICS THAT WOULD HAVE CAUSED OR CONTRIBUTED TO THE REPORTED CLINICAL OBSERVATIONS.

Description of Event or Problem · 1

--

Description of Event or Problem · 1

BOSTON SCIENTIFIC RECEIVED INFORMATION THAT THIS LEFT VENTRICULAR (LV) LEAD DISLODGED. AN INVASIVE REVISION PROCEDURE WAS PERFORMED AND THE LV LEAD WAS SUCCESSFULLY REPLACED. THERE WAS NO REPORT OF ADVERSE PATIENT EFFECTS DURING THIS PROCEDURE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
16737 EASYTRAK 2 IMPLANTABLE LEAD OJX CPI - DEL CARIBE 4542

Patients

Seq Age Sex Outcome Treatment
1 91 YR Hospitalization| L| R 4542| 4136| V173| 4469