FDA Adverse Event Malfunction Summary report: N

VITALITY 2

MDR report key: 2911396 · Received January 11, 2013

Report

Report Number
2124215-2012-16153
Event Type
Malfunction
Date Received
January 11, 2013
Date of Event
November 27, 2012
Report Date
January 9, 2013
Manufacturer
GUIDANT CRM CLONMEL IRELAND
Product Code
LWS
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

ALL AVAILABLE INFORMATION INDICATES THE DEVICE REMAINS IN SERVICE. THIS INVESTIGATION WILL BE UPDATED SHOULD FURTHER INFORMATION BE PROVIDED.

Additional Manufacturer Narrative · 1

UPON RECEIPT AT OUR POST MARKET QUALITY ASSURANCE LABORATORY, A REVIEW OF DEVICE MEMORY REVEALED THAT THE DEVICE DECLARED END OF LIFE (EOL) AFTER EXPERIENCING ONE CHARGE TIME GREATER THAN 30 SECONDS. THE OBSERVED RATE OF BATTERY USAGE WAS COMPARED TO THE EXPECTED RATE OF BATTERY USAGE; THE RESULTS INDICATED THAT THE MONITORING VOLTAGE WAS NORMAL. ALTHOUGH THE BATTERY ITSELF HAD NOT DEPLETED PREMATURELY, EOL WAS TRIGGERED EARLIER THAN EXPECTED BY EXTENDED CHARGE TIMES [, AND THE ERI TO EOL TIME PERIOD WAS SHORTENED,] DUE TO A HIGHER-THAN-TYPICAL BUILD-UP OF INTERNAL BATTERY IMPEDANCE.

Description of Event or Problem · 1

BOSTON SCIENTIFIC RECEIVED INFORMATION THAT THIS IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD) EXHIBITED ELECTIVE REPLACEMENT INDICATOR (ERI) SOONER THAN EXPECTED. BOSTON SCIENTIFIC TECHNICAL SERVICES (TS) ADVISED THIS DEVICE COULD PREMATURELY DEPLETE BASED ON MONITORING VOLTAGE AND LONGER CHARGE TIMES. THERE WERE NO ADVERSE PATIENT EFFECTS REPORTED.

Description of Event or Problem · 1

--

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
18370 VITALITY 2 IMPLANTABLE CARDIOVERTER DEFIBRILLATOR LWS GUIDANT CRM CLONMEL IRELAND T177

Patients

Seq Age Sex Outcome Treatment
1 60 YR 0181| 6936| T177| MISMATCH