FINELINE II
Report
- Report Number
- 2124215-2012-16013
- Event Type
- Injury
- Date Received
- January 11, 2013
- Date of Event
- November 28, 2012
- Report Date
- January 16, 2013
- Manufacturer
- CPI - DEL CARIBE
- Product Code
- NVN
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IN
- Reporter Occupation
- MEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE
Narratives
AS NO FURTHER INFORMATION CONCERNING THIS REPORT IS EXPECTED, OUR INVESTIGATION IS COMPLETE. THIS INVESTIGATION WILL BE UPDATED SHOULD FURTHER INFORMATION BE PROVIDED.
UPON RECEIPT AT OUR POST MARKET QUALITY ASSURANCE LABORATORY, VISUAL INSPECTION OF THE LEAD WAS PERFORMED. INSPECTION OF THE LEAD BODY AND ELECTRODE TIP FOUND A CUT IN THE INSULATION AT 205MM FROM THE TERMINAL PIN, AS A RESULT OF THE REMOVAL OF THE SUTURE SLEEVE TIE DOWN. LABORATORY TESTING DID NOT IDENTIFY ANY LEAD CHARACTERISTICS THAT WOULD HAVE RESULTED IN THE CLINICAL OBSERVATION OF DISLODGEMENT.
BOSTON SCIENTIFIC RECEIVED INFORMATION THAT THIS RIGHT ATRIAL (RA) LEAD WAS DISPLAYED NO CAPTURE AT INCREASED OUTPUTS. THE LEAD WAS FOUND TO HAVE MICRODISLODGED. A REVISION PROCEDURE WAS PERFORMED AND THE LEAD WAS ULTIMATELY EXPLANTED. NO ADVERSE PATIENT EFFECTS WERE REPORTED DURING THE PROCEDURE.
THE LEAD WAS RETURNED TO ALLOW FOR RELIABILITY ANALYSIS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 17055 | FINELINE II | IMPLANTABLE LEAD | NVN | CPI - DEL CARIBE | 4469 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 68 YR | Hospitalization| L| R | 5076| 4469| K063 |