FDA Adverse Event Malfunction Summary report: N

FLOGARD

MDR report key: 2911120 · Received January 11, 2013

Report

Report Number
1416980-2013-00911
Event Type
Malfunction
Date Received
January 11, 2013
Date of Event
December 20, 2012
Report Date
December 20, 2012
Manufacturer
BAXTER HEALTHCARE - SINGAPORE
Product Code
FRN
PMA / PMN Number
K915522
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PR
Reporter Occupation
SERVICE AND TESTING PERSONNEL

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE DEVICE HAS BEEN RECEIVED BY BAXTER, AND THE CONDITION WAS CONFIRMED BY SERVICE. THIS IS AN ANCILLARY SERVICE EVENT DISCOVERED DURING PREVENTATIVE MAINTENANCE. THE CAUSE WAS IDENTIFIED TO BE WEAR AND TEAR OF THE PUMP HEAD DOOR. TO CORRECT THIS CONDITION, THE PUMP HEAD DOOR ASSEMBLY WAS REPLACED AND THE OCCLUSION DETECTORS WERE CALIBRATED AS PER SERVICE MANUAL. IF ANY ADDITIONAL INFORMATION IS RECEIVED, A FOLLOW-UP WILL BE SENT.

Description of Event or Problem · 1

DURING PRODUCT EVALUATION BY A BAXTER SERVICE TECHNICIAN, A FLOGARD VOLUMETRIC INFUSION PUMP WAS FOUND TO HAVE PHYSICAL DAMAGE ON THE PUMP HEAD DOOR. THIS WAS NOT REPORTED BY THE CUSTOMER. THIS CONDITION HAD THE POTENTIAL TO INTERRUPT DELIVERY. THERE WAS NO PATIENT INVOLVEMENT; THEREFORE, NO PATIENT INJURY, MEDICAL INTERVENTION, OR ADVERSE REACTION IS ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
16665 FLOGARD PUMP, INFUSION FRN BAXTER HEALTHCARE - SINGAPORE

Patients

Seq Age Sex Outcome Treatment
1