FDA Adverse Event Malfunction Summary report: N

COGNIS

MDR report key: 2910875 · Received January 11, 2013

Report

Report Number
2124215-2013-00613
Event Type
Malfunction
Date Received
January 11, 2013
Date of Event
October 3, 2012
Report Date
February 17, 2017
Manufacturer
GUIDANT CRM CLONMEL IRELAND
Product Code
LWP
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AZ
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

AVAILABLE INFORMATION INDICATES THAT THIS PRODUCT REMAINS IMPLANTED AND IN SERVICE. AS NO FURTHER INFORMATION CONCERNING THIS REPORT IS EXPECTED, OUR INVESTIGATION IS COMPLETE. THIS INVESTIGATION WILL BE UPDATED SHOULD FURTHER INFORMATION BE PROVIDED.

Additional Manufacturer Narrative · 1

UPON RECEIPT AT OUR POST MARKET QUALITY ASSURANCE LABORATORY, A THOROUGH EVALUATION OF THE DEVICE WAS PERFORMED. A VISUAL INSPECTION OF THE DEVICE HEADER AND CASE NOTED NO ANOMALIES. PIN GAUGE TESTING, DESIGNED TO VERIFY PROPER PORT DIMENSIONS, WAS COMPLETED. EACH PORT MEASURED AS EXPECTED. THE DEVICE WAS THEN EXPOSED TO SIMULATED HEART LOAD CONDITIONS, AND THE DEFIBRILLATION, PACING, AND SENSING FUNCTIONS WERE TESTED. IMPEDANCE TESTING WAS COMPLETED AND ALL MEASUREMENTS WERE WITHIN NORMAL LIMITS. THE DEVICE OPERATED APPROPRIATELY WITH NO INTERRUPTIONS IN THERAPY OUTPUT AT THE RETURNED PROGRAMMED SETTINGS. A SERIES OF ELECTRICAL TESTS WAS ALSO PERFORMED, AND AGAIN, NORMAL DEVICE FUNCTION WAS OBSERVED.

Description of Event or Problem · 1

BOSTON SCIENTIFIC RECEIVED INFORMATION THAT THIS CARDIAC RESYNCHRONIZATION THERAPY DEFIBRILLATOR (CRT-D) AND IMPLANTABLE DEFIBRILLATION LEAD EXHIBITED A LOW OUT OF RANGE SHOCK IMPEDANCE MEASUREMENT. ALL SUBSEQUENT SHOCK IMPEDANCE MEASUREMENTS WERE OBSERVED TO BE STABLE AND ACCEPTABLE. NO ADVERSE PATIENT EFFECTS WERE REPORTED.

Description of Event or Problem · 1

ADDITIONAL INFORMATION WAS RECEIVED THAT THE DEVICE WAS EXPLANTED AND REPLACED 4 YEARS LATER.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
17559 COGNIS IMPLANTABLE CHF PULSE GENERATOR LWP GUIDANT CRM CLONMEL IRELAND N118

Patients

Seq Age Sex Outcome Treatment
1 76 YR (B)(4)| MISMATCH| N118| 4517