- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- MEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE
Narratives
Additional Manufacturer Narrative · 1
THE PRODUCT HAS BEEN RECEIVED AND IS CURRENTLY BEING ANALYZED. WHEN TESTING IS COMPLETE THIS REPORT WILL BE UPDATED.
Additional Manufacturer Narrative · 1
UPON RECEIPT AT OUR POST MARKET QUALITY ASSURANCE LABORATORY, VISUAL INSPECTION OF THE LEAD WAS PERFORMED. INSPECTION OF THE LEAD BODY AND ELECTRODE TIP FOUND NO ANOMALIES. LABORATORY TESTING DID NOT IDENTIFY ANY LEAD CHARACTERISTICS THAT WOULD HAVE RESULTED IN THE CLINICAL OBSERVATION OF DISLODGEMENT.
Description of Event or Problem · 1
BOSTON SCIENTIFIC RECEIVED INFORMATION THAT THIS LEFT VENTRICULAR (LV) LEAD DISLODGED. AS A RESULT, AN INVASIVE REVISION PROCEDURE WAS PERFORMED AND THE LEAD WAS SUCCESSFULLY EXPLANTED AND REPLACED. THERE WAS NO REPORT OF ADVERSE PATIENT EFFECTS DURING THIS PROCEDURE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 16414 | ACUITY | IMPLANTABLE LEAD | OJX | CPI - DEL CARIBE | 4554 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 66 YR | Hospitalization| L| R | 0292| 4554| N160| 4136| 4592 |