MDR report key: 2910722 · Received January 11, 2013

Report

Report Number
2124215-2012-16355
Event Type
Injury
Date Received
January 11, 2013
Date of Event
November 14, 2012
Report Date
December 17, 2012
Manufacturer
CPI - DEL CARIBE
Product Code
OJX
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
MEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE

Narratives

Additional Manufacturer Narrative · 1

THE PRODUCT HAS BEEN RECEIVED AND IS CURRENTLY BEING ANALYZED. WHEN TESTING IS COMPLETE THIS REPORT WILL BE UPDATED.

Additional Manufacturer Narrative · 1

UPON RECEIPT AT OUR POST MARKET QUALITY ASSURANCE LABORATORY, VISUAL INSPECTION OF THE LEAD WAS PERFORMED. INSPECTION OF THE LEAD BODY AND ELECTRODE TIP FOUND NO ANOMALIES. LABORATORY TESTING DID NOT IDENTIFY ANY LEAD CHARACTERISTICS THAT WOULD HAVE RESULTED IN THE CLINICAL OBSERVATION OF DISLODGEMENT.

Description of Event or Problem · 1

BOSTON SCIENTIFIC RECEIVED INFORMATION THAT THIS LEFT VENTRICULAR (LV) LEAD DISLODGED. AS A RESULT, AN INVASIVE REVISION PROCEDURE WAS PERFORMED AND THE LEAD WAS SUCCESSFULLY EXPLANTED AND REPLACED. THERE WAS NO REPORT OF ADVERSE PATIENT EFFECTS DURING THIS PROCEDURE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
16414 ACUITY IMPLANTABLE LEAD OJX CPI - DEL CARIBE 4554

Patients

Seq Age Sex Outcome Treatment
1 66 YR Hospitalization| L| R 0292| 4554| N160| 4136| 4592