CONTAK RENEWAL
Report
- Report Number
- 2124215-2012-15611
- Event Type
- Malfunction
- Date Received
- January 11, 2013
- Date of Event
- November 16, 2012
- Report Date
- May 27, 2015
- Manufacturer
- GUIDANT CRM CLONMEL IRELAND
- Product Code
- LWP
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- PHYSICIAN
Narratives
AS NO FURTHER INFORMATION CONCERNING THIS REPORT IS EXPECTED, OUR INVESTIGATION IS COMPLETE. THIS INVESTIGATION WILL BE UPDATED SHOULD FURTHER INFORMATION BE PROVIDED.
(B)(4). UPON RECEIPT AT OUR POST MARKET QUALITY ASSURANCE LABORATORY, A THOROUGH EVALUATION OF THE DEVICE WAS PERFORMED. A VISUAL INSPECTION OF THE DEVICE HEADER AND CASE NOTED NO ANOMALIES. THE DEVICE WAS THEN EXPOSED TO SIMULATED HEART LOAD CONDITIONS, AND THE DEFIBRILLATION, PACING, AND SENSING FUNCTIONS WERE TESTED. IMPEDANCE TESTING WAS COMPLETED AND ALL MEASUREMENTS WERE WITHIN NORMAL LIMITS. THE DEVICE OPERATED APPROPRIATELY WITH NO INTERRUPTIONS IN THERAPY OUTPUT AT THE RETURNED PROGRAMMED SETTINGS. A SERIES OF ELECTRICAL TESTS WAS ALSO PERFORMED, AND AGAIN, NORMAL DEVICE FUNCTION WAS OBSERVED.
BOSTON SCIENTIFIC RECEIVED INFORMATION THAT THE RIGHT VENTRICULAR (RV) LEAD THRESHOLD MEASUREMENTS HAVE INCREASED GRADUALLY. NO NOISE OR OUT OR RANGE IMPEDANCE MEASUREMENTS HAVE BEEN OBSERVED. AN X-RAY WAS PERFORMED AND REVEALED AN UNDERINSERTION OF THE RV LEAD. NO ADVERSE PATIENT EFFECTS WERE REPORTED.
THIS DEVICE WAS SUBSEQUENTLY EXPLANTED AND RETURNED FOR ANALYSIS. NO ADVERSE PATIENT EFFECTS WERE REPORTED.