FDA Adverse Event Injury Summary report: N

PATCH

MDR report key: 2910659 · Received January 11, 2013

Report

Report Number
2124215-2012-16514
Event Type
Injury
Date Received
January 11, 2013
Date of Event
November 16, 2012
Report Date
November 16, 2012
Manufacturer
HISTORICAL PUERTO RICO
Product Code
NHW
PMA / PMN Number
P830060
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NJ
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

THE LEADS REMAIN IMPLANTED AND SURGICALLY ABANDONED. AS NO FURTHER INFORMATION CONCERNING THIS REPORT IS EXPECTED, OUR INVESTIGATION IS COMPLETE. THIS INVESTIGATION WILL BE UPDATED SHOULD FURTHER INFORMATION BE PROVIDED.

Description of Event or Problem · 1

BOSTON SCIENTIFIC RECEIVED INFORMATION THAT OUT OF RANGE SHOCK IMPEDANCES GREATER THAN 125 OHMS WERE OBSERVED ON THE RIGHT VENTRICULAR (RV) LEAD. THIS WAS NOTED FOLLOWING THE RECENT DEVICE CHANGE OUT THAT OCCURRED IN THE PREVIOUS MONTH. THIS PATIENT HAS AN ABDOMINAL SYSTEM. THE CONCERN WAS THAT THIS MAY BE THE CAUSE OF THE OUT OF RANGE MEASUREMENTS. BOSTON SCIENTIFIC TECHNICAL SERVICES (TS) DISCUSSED PERFORMING FURTHER TESTING OR TO PERFORM A SAVE TO DISC FOR FURTHER INFORMATION. SUBSEQUENTLY ADDITIONAL INFORMATION WAS RECEIVED THAT FOUR DAYS FOLLOWING THIS REPORT, THIS SYSTEM WAS PART OF A REVISION DUE TO AN INFECTION. NO ADDITIONAL TESTING OR FOLLOW-UP WAS PERFORMED BECAUSE OF THE INFECTION. THERE WERE NO PATIENT SYMPTOMS OR ADVERSE EFFECTS RELATED TO THE OUT OF RANGE IMPEDANCE MEASUREMENTS. THE LEADS WERE ABANDONED AND REMAIN IMPLANTED. THE DEVICE WILL BE RETURNED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
17778 PATCH IMPLANTABLE LEAD NHW HISTORICAL PUERTO RICO 0041

Patients

Seq Age Sex Outcome Treatment
1 51 YR Hospitalization| L| R 1857| 1746| 6162| 6931| 4320| E141| 0041| 1600| 0010