FDA Adverse Event Malfunction Summary report: N

HOMECHOICE AUTOMATED PD SET WITH CASSETTE

MDR report key: 2910441 · Received January 11, 2013

Report

Report Number
1416980-2013-00862
Event Type
Malfunction
Date Received
January 11, 2013
Date of Event
December 27, 2012
Report Date
December 28, 2012
Manufacturer
BAXTER HEALTHCARE - MOUNTAIN HOME
Product Code
FKX
PMA / PMN Number
K923065
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
US
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THIS COMPLAINT FOR A SYSTEM ERROR 2240 (AIR IN SET) IS CONFIRMED BECAUSE THE PATIENT REPORTED DISCONNECTING FROM THE TRANSFER SET DURING THERAPY, WHICH IS USE ERROR THAT CAN CAUSE SYSTEM ERROR 2240 ALARMS. THERE WAS NO ALLEGATION REPORTED AGAINST THE BAXTER PRODUCT BY THE CUSTOMER; THEREFORE, THE SAMPLE WAS NOT REQUESTED FOR EVALUATION AND A BATCH REVIEW WAS NOT CONDUCTED. A LABELING REVIEW FOUND THE LABELING TO BE ADEQUATE FOR THE USE/USER ERROR IDENTIFIED IN THIS INCIDENT. THIS ISSUE IS BEING INVESTIGATED THROUGH CAPA.

Description of Event or Problem · 1

A CUSTOMER CONTACTED BAXTER'S TECHNICAL SERVICE CENTER REPORTING A SYSTEM ERROR (SE) 2240 (AIR IN SET) DURING DWELL 2 ON THE HOME CHOICE (HC). THE HOME PATIENT (HP) STATED HE DISCONNECTED AND RECONNECTED. THE HP STATED HE PRESSED STOP BUT WASN'T SURE HE CLOSED ALL THE CLAMPS. THE HP ALSO STATED HE RECONNECTED BUT WASN'T SURE IF HE DID THAT PRIOR TO PRESSING GO. THE TECHNICAL SERVICE REPRESENTATIVE (TSR) INFORMED THE HP THAT AIR WAS INGESTED INTO THE SYSTEM AND HE WOULD NEED TO END THE THERAPY. THE HP STATED HE WOULD WAIT TO START ANOTHER THERAPY THE FOLLOWING DAY. THE TSR ASSISTED THE HP TO END THE THERAPY AND DISCONNECT. THE TSR ADVISED THE HP TO DISPOSE OF ALL THE CURRENT SUPPLIES USED. THERE WAS PATIENT INVOLVEMENT. NO PATIENT INJURY OR MEDICAL INTERVENTION WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
18045 HOMECHOICE AUTOMATED PD SET WITH CASSETTE SYSTEM, PERITONEAL, AUTOMATIC DELIVERY FKX BAXTER HEALTHCARE - MOUNTAIN HOME

Patients

Seq Age Sex Outcome Treatment
1 45 YR HOMECHOICE