ANIMAS INSULIN INFUSION PUMP
Report
- Report Number
- 2531779-2013-00511
- Event Type
- Malfunction
- Date Received
- January 11, 2013
- Report Date
- December 27, 2012
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- K042873
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- OTHER
Narratives
THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS 08/19/2013 WITH THE FOLLOWING FINDINGS: THE KEYPAD BUTTON SYMBOLS WERE FOUND TO BE WORN. DURING TESTING, THE UP ARROW, DOWN ARROW AND OK KEYPAD BUTTONS WERE FOUND TO BE INTERMITTENTLY UNRESPONSIVE; THE CONTRAST BUTTON RESPONDED APPROPRIATELY. THERE WAS EVIDENCE OF CONTAMINATION FOUND UNDER ALL KEY CONTACTS.
THE PUMP HAS NOT BEEN RETURNED TO ANIMAS. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME.
ON (B)(6) 2012, THE PATIENT CONTACTED ANIMAS, STATING THAT THE UP ARROW KEYPAD BUTTON WAS HYPER RESPONSIVE. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 16950 | ANIMAS INSULIN INFUSION PUMP | INSULIN INFUSION PUMP | LZG | ANIMAS CORPORATION | ANM 2020 INSULIN INFUSION PUMP |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 33 YR |