FDA Adverse Event Malfunction Summary report: N

ANIMAS INSULIN INFUSION PUMP

MDR report key: 2910020 · Received January 11, 2013

Report

Report Number
2531779-2013-00512
Event Type
Malfunction
Date Received
January 11, 2013
Date of Event
December 14, 2012
Report Date
December 14, 2012
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K042873
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON (B)(4) 2012, WITH THE FOLLOWING FINDINGS: DURING A LOAD STEP ATTEMPT, THE PISTON STOPPED AT 205 UNITS, INDICATING LARGE PRIME VOLUME. INVESTIGATORS WERE ABLE TO PRIME THE PUMP AND AN OCCLUSION ALARM WAS EMITTED UPON BASAL DELIVERY ATTEMPT. THE PUMP WAS OPENED AND A FORCE SENSOR CIRCUIT COMPONENT WAS FOUND TO BE MISALIGNED ON THE PRINTED CIRCUIT BOARD. (B)(6).

Description of Event or Problem · 1

THE PUMP WAS RETURNED FOR INVESTIGATION. INVESTIGATION REVEALED A MISALIGNED FORCE SENSOR CIRCUIT COMPONENT ON THE PRINTED CIRCUIT BOARD. THIS REPORT IS MADE BASED ON RESULTS OF INVESTIGATION COMPLETED ON (B)(4) 2012.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
16933 ANIMAS INSULIN INFUSION PUMP INSULIN INFUSION PUMP LZG ANIMAS CORPORATION ANM 2020 INSULIN INFUSION PUMP

Patients

Seq Age Sex Outcome Treatment
1