EASYTRAK 2
Report
- Report Number
- 2124215-2012-15397
- Event Type
- Injury
- Date Received
- January 11, 2013
- Date of Event
- November 6, 2012
- Report Date
- December 3, 2012
- Manufacturer
- CPI - DEL CARIBE
- Product Code
- OJX
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- OK
- Reporter Occupation
- PHYSICIAN
Narratives
THE PRODUCT IS EXPECTED TO BE RETURNED FOR ANALYSIS. THIS REPORT WILL BE UPDATED UPON RETURN AND COMPLETION OF ANALYSIS.
UPON RECEIPT AT OUR POST MARKET QUALITY ASSURANCE LABORATORY, VISUAL INSPECTION OF THE LEAD CONFIRMED THAT THE CONDUCTOR COILS WERE FLATTENED AND FRACTURED APPROXIMATELY 36-38 CENTIMETERS (CM) FROM THE TERMINAL PIN. MICROSCOPIC EVALUATION INDICATED THAT THE INSULATION DAMAGE WAS CAUSED BY LOCALIZED COMPRESSIVE STRESS ON THE INSULATION SURFACE. DUE TO THE LOCATION AND THE TYPE OF DAMAGE EXHIBITED, IT WAS CONCLUDED THAT THE DAMAGE WAS CAUSED BY LEAD ENTRAPMENT IN THE CLAVICLE-FIRST RIB REGION.
BOSTON SCIENTIFIC RECEIVED INFORMATION THAT DURING A ROUTINE IN-CLINIC FOLLOW UP, THIS LEFT VENTRICULAR (LV) LEAD EXHIBITED HIGH OUT OF RANGE PACING IMPEDANCE MEASUREMENTS AND NO PACING. THE PATIENT EXPERIENCED SYMPTOMS; INCLUDING SHORTNESS OF BREATH. AN X-RAY WAS PERFORMED. THE LEAD WAS OBSERVED TO HAVE FRACTURED. AN INVASIVE PROCEDURE WAS PERFORMED TWO DAYS LATER. THE LEAD WAS SUCCESSFULLY EXPLANTED. IT WAS REPORTED THAT THE BRANCH COLLAPSED ON ITSELF UPON REMOVAL OF THE LEAD. AN ATTEMPT WAS MADE TO IMPLANT A NEW LV LEAD WITH NO SUCCESS. THE PROCEDURE WAS ABORTED AND ANOTHER PROCEDURE WILL LIKELY BE SCHEDULED FOR A DATE IN THE FUTURE. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 17638 | EASYTRAK 2 | IMPLANTABLE LEAD | OJX | CPI - DEL CARIBE | 4542 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 66 YR | Hospitalization| L| R | 0185| 4136| 4542| E110| N119 |