FDA Adverse Event Malfunction Summary report: N

OT VERIO IQ METER

MDR report key: 2905794 · Received January 9, 2013

Report

Report Number
3008382007-2013-00533
Event Type
Malfunction
Date Received
January 9, 2013
Report Date
December 18, 2012
Manufacturer
LIFESCAN EUROPE, A DIVISION OF CILAG GMBH INTL
Product Code
NBW
PMA / PMN Number
K110637
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

FOLLOW-UP # 1 (03/07/2013)-DEVICE EVALUATION: THE PRODUCTS INVOLVED WITH THIS COMPLAINT HAVE BEEN RETURNED ON (B)(4) 2013 AND EVALUATED BY PRODUCT ANALYSIS (PA) ON (B)(4) 2013 WITH THE FOLLOWING FINDINGS: THE METER PASSED TESTING WITH NO FAULTS FOUND. THE METER WAS FOUND TO FUNCTION PROPERLY. THE USB CABLE AND ADAPTER BOTH ALSO PASSED TESTING WITH NO FAULTS FOUND. IF LIFESCAN OBTAINS ADDITIONAL INFORMATION REGARDING THIS COMPLAINT, A FOLLOW UP REPORT WILL BE SUBMITTED. AT THIS TIME, LIFESCAN CONSIDERS THIS MATTER CLOSED.

Description of Event or Problem · 1

ON (B)(6) 2012, THE REPORTER CONTACTED LIFESCAN USA, ALLEGING BATTERY INDICATOR. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
10683 OT VERIO IQ METER GLUCOSE MONITORING SYS/KIT NBW LIFESCAN EUROPE, A DIVISION OF CILAG GMBH INTL 3237365

Patients

Seq Age Sex Outcome Treatment
1 65 YR