FDA Adverse Event Malfunction Summary report: N

OT VERIO IQ METER

MDR report key: 2905777 · Received January 9, 2013

Report

Report Number
3008382007-2013-00477
Event Type
Malfunction
Date Received
January 9, 2013
Report Date
December 15, 2012
Manufacturer
LIFESCAN EUROPE, A DIVISION OF CILAG GMBH INTL
Product Code
NBW
PMA / PMN Number
K110637
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE SUBJECT METER AND TEST STRIPS HAVE BEEN RETURNED TO LIFESCAN FOR EVALUATION. THE EVALUATION OF THE PRODUCTS HAVE NOT BEEN COMPLETED; THEREFORE, NO CONCLUSIONS CAN BE DRAWN AT THIS TIME. ONCE THE EVALUATION IS COMPLETED, A SUPPLEMENTAL REPORT WILL BE FILED.

Additional Manufacturer Narrative · 1

(B)(4): THE REPORTED COMPLAINT WAS OBSERVED ON THE METER'S ERROR LOG; HOWEVER, PA WAS UNABLE TO REPRODUCE FAILURE IN METER TESTING. THE TEST STRIPS INVOLVED WITH THIS COMPLAINT WERE EVALUATED AND ER4 WAS OBSERVED WITH CONTROL SOLUTION TESTING. IF LIFESCAN OBTAINS ADDITIONAL INFORMATION REGARDING THIS COMPLAINT, A FOLLOW UP REPORT WILL BE SUBMITTED. AT THIS TIME, LIFESCAN CONSIDERS THIS MATTER CLOSED.

Description of Event or Problem · 1

ON (B)(6) 2012, THE REPORTER CONTACTED LIFESCAN (LFS) IN THE USA ALLEGING THE METER WAS DISPLAYING AN ERROR 4 MESSAGE. THIS COMPLAINT IS BEING REPORTED BECAUSE THE ALLEGED ISSUE WAS NOT RESOLVE WITH TROUBLESHOOTING DONE BY THE CUSTOMER CARE ADVOCATE (CCA). THERE WAS NO INDICATION THAT THE LFS PRODUCT CAUSED OR CONTRIBUTED TO A ADVERSE EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
12421 OT VERIO IQ METER GLUCOSE MONITORING SYS/KIT NBW LIFESCAN EUROPE, A DIVISION OF CILAG GMBH INTL