COULTER® HMX HEMATOLOGY ANALYZER WITH AUTOLOADER
Report
- Report Number
- 1061932-2013-00028
- Event Type
- Malfunction
- Date Received
- January 9, 2013
- Date of Event
- December 11, 2012
- Report Date
- December 12, 2012
- Manufacturer
- BECKMAN COULTER, INC.
- Product Code
- GKZ
- PMA / PMN Number
- K010765
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- SC, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
THE FIELD SERVICE ENGINEER (FSE) NOTED THE BLOOD SAMPLING VALVE (BSV) WAS NOT ROTATING DUE TO A DEFECTIVE BSV ACTUATOR. THE FSE REPLACED THE ACTUATOR, BSV, TUBING, AND SOLENOIDS 16, 17, 18, AND 19. THE FSE VERIFIED THE REPAIR PER ESTABLISHED PROCEDURES. THE INSTRUMENT CONFORMED TO THE MANUFACTURER'S PUBLISHED PERFORMANCE SPECIFICATIONS.
THE CUSTOMER REPORTED DISCORDANT HEMOGLOBIN AND HEMATOCRIT RESULTS, FOR SEVERAL PATIENTS, INVOLVING THE COULTER HMX HEMATOLOGY ANALYZER WITH AUTOLOADER. THE PATIENT SAMPLES WERE ANALYZED BY A REFERENCE LABORATORY AND RECOVERED HEMOGLOBIN AND HEMATOCRIT RESULTS THAT WERE CONSIDERED CORRECT. THE ERRONEOUS RESULTS WERE NOT RELEASED OUT OF THE LABORATORY. THERE WAS NO PATIENT CONSEQUENCE ASSOCIATED WITH THIS EVENT. CONTROLS WERE WITHIN SPECIFICATION AT THE TIME OF THE EVENT. BECKMAN COULTER FIELD SERVICE ENGINEER (FSE) WAS DISPATCHED TO EVALUATE THE INSTRUMENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 12694 | COULTER® HMX HEMATOLOGY ANALYZER WITH AUTOLOADER | COUNTER, DIFFERENTIAL CELL | GKZ | BECKMAN COULTER, INC. | NA | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |