FDA Adverse Event Injury Summary report: N

PARADIGM REAL-TIME INSULIN INFUSION PUMP

MDR report key: 2905643 · Received January 9, 2013

Report

Report Number
3004209178-2013-90164
Event Type
Injury
Date Received
January 9, 2013
Date of Event
December 21, 2012
Report Date
December 22, 2012
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OYC
PMA / PMN Number
P980022
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
CA
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

THE INSULIN PUMP WAS RECEIVED WITH OPERATING CURRENTS WITHIN SPECIFICATION. THE INSULIN PUMP PASSED THE REWIND, BASIC OCCLUSION TEST, PRIME AND DISPLACEMENT. HOWEVER, THE INSULIN PUMP WAS RECEIVED STUCK IN THE MOTOR ERROR LOOP DURING BOLUS AND BASAL DELIVERY. ERROR ALARM CONFIRMED IN THE HISTORY FILE. THE MOTOR WAS TESTED OUTSIDE THE DEVICE AND PASSED. THE MOTOR MAY HAVE HAD AN INTERMITTENT FAILURE THAT WAS NOT DETECTED DURING OUR TESTING. UNABLE TO PERFORM ERROR TEST, OCCLUSION TEST, EXCESSIVE NO DELIVERY TEST DUE TO MOTOR ERROR ALARMS.

Additional Manufacturer Narrative · 1

FAILURE ANALYSIS REVEALED THAT THE INSULIN PUMP WAS RECEIVED WITH OPERATING CURRENTS WITHIN SPECIFICATIONS. THE DEVICE PASSED THE REWIND, BASIC OCCLUSION, PRIME, AND DISPLACEMENT TEST. HOWEVER, THE INSULIN PUMP WAS RECEIVED STUCK IN THE MOTOR LOOP DURING BOLUS AND BASAL DELIVERY. THE ERROR ALARMS WERE CONFIRMED IN THE ALARM HISTORY. THE MOTOR WAS TESTED OUTSIDE THE UNIT AND PASSED. THE MOTOR MAY HAVE AN INTERMITTENT FAILURE THAT WAS NOT DETECTED DURING OUR TESTING. FURTHER TESTING COULD NOT BE PERFORMED DUE TO THE MOTOR ERROR ALARMS.

Description of Event or Problem · 1

THE CUSTOMER REPORTED BEING IN THE EMERGENCY ROOM DUE TO DIABETES KETOACIDOSIS AND HIGH BLOOD GLUCOSE OF 506MG/DL. THE CALLER STATED THAT HE RECEIVED MOTOR ERROR ALARMS AND OTHER ERROR ALARMS. ADVISED THE CUSTOMER THAT THE INSULIN PUMP WOULD BE REPLACED AND REVERT TO BACK UP PLAN. NO FURTHER INFORMATION WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
12528 PARADIGM REAL-TIME INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC PUERTO RICO OPERATIONS CO. MMT-722NAS

Patients

Seq Age Sex Outcome Treatment
1 35 YR Hospitalization