FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME INSULIN INFUSION PUMP

MDR report key: 2905608 · Received January 9, 2013

Report

Report Number
3004209178-2013-90183
Event Type
Malfunction
Date Received
January 9, 2013
Date of Event
October 22, 2012
Report Date
January 3, 2013
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TN
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

FAILURE ANALYSIS REVEALED THAT THE INSULIN PUMP ALARMED DURING THE BASIC OCCLUSION TEST, AND THE DEVICE WAS UNABLE TO PRIME DURING REWIND AS A RESULT OF A LOOSE DRIVE SUPPORT DISK.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT THE INSULIN PUMP ALARMED DURING THE MANUAL PRIME PROCESS. ADVISED THE CUSTOMER THAT THE DEVICE WOULD BE REPLACED. THE BLOOD GLUCOSE READING WAS 328MG/DL. NO FURTHER INFORMATION WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
11433 PARADIGM REAL-TIME INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC PUERTO RICO OPERATIONS CO. MMT-722NAS

Patients

Seq Age Sex Outcome Treatment
1 49 YR