FDA Adverse Event Malfunction Summary report: N

FLOGARD

MDR report key: 2905596 · Received January 9, 2013

Report

Report Number
1416980-2013-00704
Event Type
Malfunction
Date Received
January 9, 2013
Date of Event
November 16, 2012
Report Date
November 16, 2012
Manufacturer
SHARP CORPORATION OSAKA JAPAN
Product Code
FRN
PMA / PMN Number
K915522
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MX
Reporter Occupation
SERVICE AND TESTING PERSONNEL

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE DEVICE WAS REPAIRED ON-SITE AT THE CUSTOMER FACILITY BY A BAXTER FIELD SERVICE TECHNICIAN, AND THE CONDITION WAS CONFIRMED. THIS IS AN ANCILLARY SERVICE EVENT DISCOVERED DURING PREVENTATIVE MAINTENANCE. THE CAUSE WAS DETERMINED TO BE A DEFECTIVE BATTERY. TO CORRECT THE CONDITION, THE MAIN BATTERY WAS REPLACED. IF ANY ADDITIONAL INFORMATION IS RECEIVED, A FOLLOW-UP WILL BE SENT.

Description of Event or Problem · 1

DURING PRODUCT EVALUATION BY A BAXTER SERVICE TECHNICIAN, A FLO-GARD INFUSION PUMP WAS FOUND TO HAVE EXPERIENCED FAILURE CODE F-94. THIS CONDITION HAD THE POTENTIAL TO INTERRUPT DELIVERY. THIS WAS NOT REPORTED BY THE CUSTOMER. THERE WAS NO PATIENT INVOLVEMENT; THEREFORE, NO PATIENT INJURY, MEDICAL INTERVENTION, OR ADVERSE REACTION IS ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
11429 FLOGARD PUMP, INFUSION FRN SHARP CORPORATION OSAKA JAPAN

Patients

Seq Age Sex Outcome Treatment
1