XIENCE PRIME LL EVEROLIMUS ELUTING CORONARY STENT SYSTEM
Report
- Report Number
- 2024168-2013-00185
- Event Type
- Malfunction
- Date Received
- January 9, 2013
- Date of Event
- December 17, 2012
- Report Date
- December 18, 2012
- Manufacturer
- AV-TEMECULA-CT
- Product Code
- NIQ
- PMA / PMN Number
- P110019
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- LE
- Reporter Occupation
- HEALTH PROFESSIONAL
Narratives
(B)(4). DURING PROCESSING OF THIS COMPLAINT, ATTEMPTS WERE MADE TO OBTAIN COMPLETE EVENT, PATIENT AND DEVICE INFORMATION. THE DEVICE IS EXPECTED TO BE RETURNED FOR INVESTIGATION. IT HAS NOT YET BEEN RECEIVED. A FOLLOW-UP REPORT WILL BE SUBMITTED WITH ALL RELEVANT INFORMATION.
(B)(4). EVALUATION SUMMARY: THE DEVICE WAS RETURNED FOR EVALUATION. THE REPORTED SHAFT SEPARATION WAS CONFIRMED AS ANALYSIS OF THE DEVICE REVEALED THAT THE HYPOTUBE HAD SEPARATED AT THE DISTAL END OF THE STRAIN RELIEF TUBING. BASED ON VISUAL INSPECTION OF THE RETURNED DEVICE, THERE IS NO INDICATION OF A PRODUCT DEFICIENCY. A REVIEW OF THE LOT HISTORY RECORD REVEALED NO NON-CONFORMANCES THAT WOULD HAVE CONTRIBUTED TO THE REPORTED EVENT. RESULTS OF THE QUERY OF SIMILAR INCIDENTS IN THE COMPLAINT-HANDLING DATABASE FROM THIS LOT INDICATED DID NOT INDICATE A MANUFACTURING ISSUE. BASED ON THE INFORMATION REVIEWED, THERE IS NO INDICATION OF A PRODUCT DEFICIENCY.
IT WAS REPORTED THAT WHILE UNPACKING, THE HUB OF THE XIENCE PRIME STENT DELIVERY SYSTEM WAS FOUND DETACHED FROM THE (CATHETER) SHAFT. THERE WAS NO PATIENT INVOLVEMENT AND THE DEVICE WAS NOT USED IN THE PROCEDURE. THERE WAS NO CLINICALLY SIGNIFICANT DELAY IN THE PROCEDURE. NO ADDITIONAL INFORMATION WAS PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 11934 | XIENCE PRIME LL EVEROLIMUS ELUTING CORONARY STENT SYSTEM | DRUG ELUTING CORONARY STENT SYSTEM | NIQ | AV-TEMECULA-CT | 2062141 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |