FDA Adverse Event Malfunction Summary report: N

XIENCE PRIME LL EVEROLIMUS ELUTING CORONARY STENT SYSTEM

MDR report key: 2905201 · Received January 9, 2013

Report

Report Number
2024168-2013-00185
Event Type
Malfunction
Date Received
January 9, 2013
Date of Event
December 17, 2012
Report Date
December 18, 2012
Manufacturer
AV-TEMECULA-CT
Product Code
NIQ
PMA / PMN Number
P110019
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
LE
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

(B)(4). DURING PROCESSING OF THIS COMPLAINT, ATTEMPTS WERE MADE TO OBTAIN COMPLETE EVENT, PATIENT AND DEVICE INFORMATION. THE DEVICE IS EXPECTED TO BE RETURNED FOR INVESTIGATION. IT HAS NOT YET BEEN RECEIVED. A FOLLOW-UP REPORT WILL BE SUBMITTED WITH ALL RELEVANT INFORMATION.

Additional Manufacturer Narrative · 1

(B)(4). EVALUATION SUMMARY: THE DEVICE WAS RETURNED FOR EVALUATION. THE REPORTED SHAFT SEPARATION WAS CONFIRMED AS ANALYSIS OF THE DEVICE REVEALED THAT THE HYPOTUBE HAD SEPARATED AT THE DISTAL END OF THE STRAIN RELIEF TUBING. BASED ON VISUAL INSPECTION OF THE RETURNED DEVICE, THERE IS NO INDICATION OF A PRODUCT DEFICIENCY. A REVIEW OF THE LOT HISTORY RECORD REVEALED NO NON-CONFORMANCES THAT WOULD HAVE CONTRIBUTED TO THE REPORTED EVENT. RESULTS OF THE QUERY OF SIMILAR INCIDENTS IN THE COMPLAINT-HANDLING DATABASE FROM THIS LOT INDICATED DID NOT INDICATE A MANUFACTURING ISSUE. BASED ON THE INFORMATION REVIEWED, THERE IS NO INDICATION OF A PRODUCT DEFICIENCY.

Description of Event or Problem · 1

IT WAS REPORTED THAT WHILE UNPACKING, THE HUB OF THE XIENCE PRIME STENT DELIVERY SYSTEM WAS FOUND DETACHED FROM THE (CATHETER) SHAFT. THERE WAS NO PATIENT INVOLVEMENT AND THE DEVICE WAS NOT USED IN THE PROCEDURE. THERE WAS NO CLINICALLY SIGNIFICANT DELAY IN THE PROCEDURE. NO ADDITIONAL INFORMATION WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
11934 XIENCE PRIME LL EVEROLIMUS ELUTING CORONARY STENT SYSTEM DRUG ELUTING CORONARY STENT SYSTEM NIQ AV-TEMECULA-CT 2062141

Patients

Seq Age Sex Outcome Treatment
1