FDA Adverse Event
Malfunction
Summary report: N
QUICKSITE LV
MDR report key: 2905193
·
Received January 9, 2013
Report
- Report Number
- 2017865-2013-00470
- Event Type
- Malfunction
- Date Received
- January 9, 2013
- Date of Event
- November 20, 2012
- Manufacturer
- ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION
- Product Code
- OJX
- PMA / PMN Number
- P030054
- Removal / Correction Number
- Z1482
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
Additional Manufacturer Narrative · 1
ALL INFORMATION PROVIDED BY MANUFACTURER, NO MEDWATCH FORM WAS RECEIVED. (B)(4) - NO COMPLAINT WAS RECEIVED WITH RETURN OF THE DEVICE. FAILURE (EVENT) OBSERVED DURING ANALYSIS. FINAL ANALYSIS FOUND THAT THE INSULATION WAS ABRADED BETWEEN 3.2 CM AND 3.5 CM FROM THE DISTAL TIP. ABRASIONS ARE CONSISTENT WITH CONSTANT FRICTION WITH ANOTHER IMPLANTABLE DEVICE.
Description of Event or Problem · 1
THIS REPORT IS TO ADVISE OF AN EVENT OBSERVED DURING ANALYSIS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 10703 | QUICKSITE LV | PERMANENT PACEMAKER ELECTRODE | OJX | ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION | 1056T/86 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |