FDA Adverse Event Malfunction Summary report: N

QUICKSITE LV

MDR report key: 2905193 · Received January 9, 2013

Report

Report Number
2017865-2013-00470
Event Type
Malfunction
Date Received
January 9, 2013
Date of Event
November 20, 2012
Manufacturer
ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION
Product Code
OJX
PMA / PMN Number
P030054
Removal / Correction Number
Z1482
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

ALL INFORMATION PROVIDED BY MANUFACTURER, NO MEDWATCH FORM WAS RECEIVED. (B)(4) - NO COMPLAINT WAS RECEIVED WITH RETURN OF THE DEVICE. FAILURE (EVENT) OBSERVED DURING ANALYSIS. FINAL ANALYSIS FOUND THAT THE INSULATION WAS ABRADED BETWEEN 3.2 CM AND 3.5 CM FROM THE DISTAL TIP. ABRASIONS ARE CONSISTENT WITH CONSTANT FRICTION WITH ANOTHER IMPLANTABLE DEVICE.

Description of Event or Problem · 1

THIS REPORT IS TO ADVISE OF AN EVENT OBSERVED DURING ANALYSIS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
10703 QUICKSITE LV PERMANENT PACEMAKER ELECTRODE OJX ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION 1056T/86 NA