FDA Adverse Event
Malfunction
Summary report: N
ACCENT DR RF
MDR report key: 2905192
·
Received January 9, 2013
Report
- Report Number
- 2017865-2013-00829
- Event Type
- Malfunction
- Date Received
- January 9, 2013
- Date of Event
- October 31, 2012
- Manufacturer
- ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION
- Product Code
- DXY
- PMA / PMN Number
- NA
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
Additional Manufacturer Narrative · 1
RELIABILITY LABORATORY TECHNICIAN, NOT APPLICABLE. - ST. JUDE MEDICAL (B)(4) RELIABILITY LABORATORY. NO COMPLAINT WAS RECEIVED WITH RETURN OF THE DEVICE. FAILURE (EVENT) OBSERVED DURING ANALYSIS. FINAL ANALYSIS FOUND THAT THE RADIO FREQUENCY MODULE EXHIBITED LOSS OF COMMUNICATION.
Description of Event or Problem · 1
THIS REPORT IS TO ADVISE OF AN EVENT OBSERVED DURING ANALYSIS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 11931 | ACCENT DR RF | IMPLANTABLE PACEMAKER PULSE GENERATOR | DXY | ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION | PM2210 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |