FDA Adverse Event
Malfunction
Summary report: N
MARYLAND BIPOLAR FORCEPS INSTRUMENT
MDR report key: 2905151
·
Received January 9, 2013
Report
- Report Number
- 2955842-2013-00105
- Event Type
- Malfunction
- Date Received
- January 9, 2013
- Date of Event
- December 11, 2012
- Report Date
- December 12, 2012
- Manufacturer
- INTUITIVE SURGICAL,INC.
- Product Code
- NAY
- PMA / PMN Number
- K050369
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- KS
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
THE INSTRUMENT WAS RETURNED AND EVALUATED. PER THE CUSTOMER REPORTED COMPLAINT, ENGINEERING OBSERVED THAT THE PITCH CABLE IS BROKEN AT THE DISTAL CLEVIS HUB. THE CABLE SEGMENT THAT CONTAINS THE CRIMP REMAINS IN THE CLEVIS. THE CLEVIS DOES NOT EXHIBIT EXCESSIVE DAMAGE. THE CUSTOMER REPORTED COMPLAINT DOES NOT ITSELF CONSTITUTE A MDR REPORTABLE EVENT; HOWEVER; THE REPORTED MALFUNCTION IF TO RECUR COULD CAUSE OR CONTRIBUTE TO AN ADVERSE EVENT.
Description of Event or Problem · 1
IT WAS REPORTED THAT DURING REPROCESSING OF THE MARYLAND BIPOLAR FORCEPS INSTRUMENT IT WAS OBSERVED THAT THE WIRE ON THE INSTRUMENT WAS BROKEN. THERE WERE NO MISSING OR FALLEN PIECES REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 12009 | MARYLAND BIPOLAR FORCEPS INSTRUMENT | ENDOSCOPIC ELECTROSURGICAL INSTRUMENT | NAY | INTUITIVE SURGICAL,INC. | 420172-07 | M10111007 737 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | DA VINCI SI SYSTEM INSTRUMENTS, ACCESSORIES & ESU |