FDA Adverse Event Malfunction Summary report: N

TENDRIL ST

MDR report key: 2905085 · Received January 9, 2013

Report

Report Number
2017865-2013-00673
Event Type
Malfunction
Date Received
January 9, 2013
Date of Event
October 31, 2012
Manufacturer
ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION
Product Code
NVN
PMA / PMN Number
P960013
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

ALL INFORMATION PROVIDED BY MANUFACTURER, NO MEDWATCH FORM WAS RECEIVED. (B)(4)- NO COMPLAINT WAS RECEIVED WITH RETURN OF THE DEVICE. FAILURE (EVENT) OBSERVED DURING ANALYSIS. FINAL ANALYSIS FOUND THAT A PARTIAL LEAD WAS RETURNED. THE OUTER INSULATION WAS ABRADED AT 5.7 CM TO 6.0 CM FROM THE CONNECTOR PIN WHICH EXPOSED THE OUTER COIL. ABRASIONS ARE INDICATIVE OF EXPOSURE TO CONSTANT FRICTION WITH ANOTHER IMPLANTABLE DEVICE.

Description of Event or Problem · 1

THIS REPORT IS TO ADVISE OF AN EVENT OBSERVED DURING ANALYSIS.