FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME INSULIN INFUSION PUMP

MDR report key: 2905084 · Received January 9, 2013

Report

Report Number
2032227-2013-00074
Event Type
Malfunction
Date Received
January 9, 2013
Date of Event
December 19, 2013
Report Date
December 19, 2013
Manufacturer
MEDTRONIC MINIMED
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
SF
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE INSULIN PUMP ALARMED MOTOR ERROR DURING THE BASIC OCCLUSION TEST DUE TO A LOOSE MOTOR SUPPORT DISK.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE PLUNGER DOESN'T STOP WHEN PERFORMING THE MANUAL PRIME. NOTHING FURTHER WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
11820 PARADIGM REAL-TIME INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC MINIMED MMT-722WWL

Patients

Seq Age Sex Outcome Treatment
1