FDA Adverse Event
Malfunction
Summary report: N
PASSIVE PLUS DX
MDR report key: 2905076
·
Received January 9, 2013
Report
- Report Number
- 2017865-2013-00503
- Event Type
- Malfunction
- Date Received
- January 9, 2013
- Date of Event
- October 7, 2012
- Manufacturer
- ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION
- Product Code
- NVN
- PMA / PMN Number
- P960030
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
Additional Manufacturer Narrative · 1
RELIABILITY LABORATORY TECHNICIAN, NOT APPLICABLE. - ST. JUDE MEDICAL (B)(4) RELIABILITY LABORATORY. NO COMPLAINT WAS RECEIVED WITH RETURN OF THE DEVICE. FAILURE (EVENT) OBSERVED DURING ANALYSIS. FINAL ANALYSIS FOUND THAT A PARTIAL LEAD WAS RETURNED. AN OUTER INSULATION ABRASION WAS NOTED BETWEEN 30 CM TO 31 CM FROM THE CONNECTOR PIN. ABRASIONS ARE CONSISTENT WITH CONSTANT FRICTION WITH ANOTHER IMPLANTABLE DEVICE.
Description of Event or Problem · 1
THIS REPORT IS TO ADVISE OF AN EVENT OBSERVED DURING ANALYSIS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 11127 | PASSIVE PLUS DX | PERMANENT PACEMAKER ELECTRODE | NVN | ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION | 1346T/58 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |