FDA Adverse Event
Malfunction
Summary report: N
PARADIGM REAL-TIME INSULIN INFUSION PUMP
MDR report key: 2905062
·
Received January 9, 2013
Report
- Report Number
- 2032227-2013-00071
- Event Type
- Malfunction
- Date Received
- January 9, 2013
- Date of Event
- December 20, 2013
- Report Date
- December 20, 2013
- Manufacturer
- MEDTRONIC MINIMED
- Product Code
- OYC
- PMA / PMN Number
- P980022
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TN
- Reporter Occupation
- PATIENT FAMILY MEMBER OR FRIEND
Narratives
Additional Manufacturer Narrative · 1
THE INSULIN PUMP ALARMED DURING THE BASIC OCCLUSION TEST DUE TO A LOOSE DRIVE SUPPORT DISK. THE INSULIN PUMP HAD A MISSING END CAP STICKER.
Description of Event or Problem · 1
THE CUSTOMER REPORTED AN ALARM AND INSULIN SQUIRTING OUT DURING THE MANUAL PRIME. ADVISED THE CUSTOMER THAT THE INSULIN PUMP WOULD BE REPLACED. NOTHING FURTHER WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 11624 | PARADIGM REAL-TIME INSULIN INFUSION PUMP | INSULIN INFUSION PUMP / SENSOR AUGMENTED | OYC | MEDTRONIC MINIMED | MMT-522LNAB |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 10 YR |