FDA Adverse Event Malfunction Summary report: N

ACCENT DR RF

MDR report key: 2905027 · Received January 9, 2013

Report

Report Number
2017865-2013-00830
Event Type
Malfunction
Date Received
January 9, 2013
Date of Event
August 21, 2012
Manufacturer
ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION
Product Code
DXY
PMA / PMN Number
NA
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

RELIABILITY LABORATORY TECHNICIAN, NOT APPLICABLE.- ST. JUDE MEDICAL (B)(4) RELIABILITY LABORATORY. NO COMPLAINT WAS RECEIVED WITH RETURN OF THE DEVICE. FAILURE (EVENT) OBSERVED DURING ANALYSIS. FINAL ANALYSIS FOUND A SOLDER BRIDGE ANOMALY WHICH DRAINED THE BATTERY PREMATURELY.

Additional Manufacturer Narrative · 1

ALL INFORMATION PROVIDED BY MANUFACTURER, NO MEDWATCH FORM WAS RECEIVED. THIS HISTORICAL COMPLAINT IS BEING FILED AS PART OF A RETROSPECTIVE REVIEW OF COMPLAINT FILES IN RESPONSE TO A RECENT FDA INSPECTION. THERE IS NO CHANGE TO THE ACTUAL PERFORMANCE OF THE PRODUCT AND THIS REPORT ONLY REPRESENTS AN ENHANCEMENT TO THE REPORTING CRITERIA GOING FORWARD.

Description of Event or Problem · 1

THIS REPORT IS TO ADVISE OF AN EVENT OBSERVED DURING ANALYSIS.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING IMPLANT, THE DEVICE WAS FOUND IN BACKUP MODE. THE DEVICE WAS NOT IMPLANTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
12636 ACCENT DR RF IMPLANTABLE PACEMAKER PULSE GENERATOR DXY ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION PM2210 NA

Patients

Seq Age Sex Outcome Treatment
1 77 YR