ACCENT DR RF
Report
- Report Number
- 2017865-2013-00830
- Event Type
- Malfunction
- Date Received
- January 9, 2013
- Date of Event
- August 21, 2012
- Manufacturer
- ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION
- Product Code
- DXY
- PMA / PMN Number
- NA
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
RELIABILITY LABORATORY TECHNICIAN, NOT APPLICABLE.- ST. JUDE MEDICAL (B)(4) RELIABILITY LABORATORY. NO COMPLAINT WAS RECEIVED WITH RETURN OF THE DEVICE. FAILURE (EVENT) OBSERVED DURING ANALYSIS. FINAL ANALYSIS FOUND A SOLDER BRIDGE ANOMALY WHICH DRAINED THE BATTERY PREMATURELY.
ALL INFORMATION PROVIDED BY MANUFACTURER, NO MEDWATCH FORM WAS RECEIVED. THIS HISTORICAL COMPLAINT IS BEING FILED AS PART OF A RETROSPECTIVE REVIEW OF COMPLAINT FILES IN RESPONSE TO A RECENT FDA INSPECTION. THERE IS NO CHANGE TO THE ACTUAL PERFORMANCE OF THE PRODUCT AND THIS REPORT ONLY REPRESENTS AN ENHANCEMENT TO THE REPORTING CRITERIA GOING FORWARD.
THIS REPORT IS TO ADVISE OF AN EVENT OBSERVED DURING ANALYSIS.
IT WAS REPORTED THAT DURING IMPLANT, THE DEVICE WAS FOUND IN BACKUP MODE. THE DEVICE WAS NOT IMPLANTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 12636 | ACCENT DR RF | IMPLANTABLE PACEMAKER PULSE GENERATOR | DXY | ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION | PM2210 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 77 YR |